Enhanced calibration method for wall-mounted RGSC cameras: Mitigating baseline drift caused by CT couch sagging.
Article in PloS one, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe drift signal due to CT couch sagging introduces intrinsic noise into respiratory signals, impairing the accuracy of tumor motion management. The purpose of this work is to present an enhanced calibration method to reduce calibration uncertainty and mitigate baseline drift caused by table sagging for the wall-mounted Respiratory Gating for Scanner (RGSC) camera.
methodsA weight of approximately 70 kg, simulating a patient's weight, was distributed on the CT table. The external ceiling laser light was adjusted laterally by ±10 cm to align three reflector blocks sequentially with the laser at predefined positions, ensuring accurate placement of the blocks at their corresponding positions. The blocks were then moved to the internal laser plane using the CT console. Subsequently, calibration measurements were performed at nine points at the combination of three lateral positions (the CT isocenter and ±10 cm laterally from the isocenter) and three longitudinal positions (the CT isocenter and ±15 cm longitudinally from the isocenter), by occluding the other two blocks and moving the couch longitudinally. For comparison, the Varian calibration method was also implemented.
resultsThe block positioning uncertainty was reduced from the millimeter level to the sub-millimeter level. For a typical 40 cm scan length of DIBH, the residual baseline drift was significantly (p-value<0.001) mitigated from 2.84 ± 0.22 mm to 0.64 ± 0.06 mm.
conclusionThe proposed calibration method provides a robust solution to minimize block positioning uncertainty and reduce baseline drift caused by CT couch sagging, enhancing the repeatability and accuracy of the wall-mounted camera calibration. Its versatility for wall- and ceiling-mounted cameras further expands its potential clinical utility.