ArticleSurgical endoscopy2025
Evaluation of automatic staplers for renal artery division in laparoscopic donor nephrectomy: impact of vessel diameter on hemostatic outcomes.
Article in Surgical endoscopy, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundLaparoscopic donor nephrectomy requires precise renal artery management for optimal outcomes. This study evaluates the performance and safety of automatic staplers in dividing renal arteries of various diameters.
methodsExperimental tests were conducted using porcine arteries (3, 5, and 7 mm) to compare double- and triple-row staplers in terms of stapling completeness and pressure resistance. Clinically, 371 renal artery resections from 276 donors were analyzed to compare hemostatic outcomes between stapler types, grouping arteries according to diameter (≤ 3, 4-6, and ≥ 7 mm).
resultsIn 7 mm vessels, both stapler types achieved complete stapling with consistent pressure resistance up to 750 mm Hg. In 5 mm vessels, both types showed comparable results, although pressure resistance was slightly higher with the double-row stapler. In 3 mm vessels, incomplete stapling was common and double-row staplers achieved a maximum pressure resistance of 700 mm Hg. Clinically, hemostatic conditions at the artery stump did not significantly differ between the stapler types for any vessel size group; however, for vessels ≥ 7 mm, the incidence of oozing, pulsatile bleeding, or bleeding requiring additional clips was higher with the triple-row stapler. The incidence of bleeding was significantly higher in vessels ≤ 3 mm (p = 0.01).
conclusionAutomatic staplers are effective for renal artery management in laparoscopic donor nephrectomy; however, when used on smaller arteries, supplemental hemostatic measures may be required. Advancing stapler technology and careful device selection are crucial for improving surgical outcomes.