Comparison of a guidewire with an uninsulated tip versus dedicated radiofrequency wire with discrete electrode for energy-based transseptal puncture: a pre-clinical study.
Article in Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Evolution of Radiofrequency Transseptal Puncture Devices.Medical devices (Auckland, N.Z.) · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundDedicated radiofrequency (RF) needles and wires for transseptal puncture (TSP) achieve better outcomes vs. electrified open-ended needles and guidewires due to optimized electrode design and energy delivery. This study benchmarked TSP performance between the dedicated VersaCross wire system (VC; Boston Scientific) and an electrified guidewire with an alternative electrode configuration similar to commercially available devices.
methodsA 0.025″ guidewire (Cordis) was modified to mimic the 15 cm 0.025″ distal uninsulated wire length of the HOTWIRE (Atraverse) device and electrified using a ValleyLab generator (EG; 30-50 W, 1 s and 300 ms). EG was extended 0-3.5 mm from the dilator tip to mimic clinical use. RF puncture using VC and EG was performed on ex vivo porcine septa (n = 18) and in vivo swine heart (n = 3), to compare TSP performance, thermal damage and energy output measured using an oscilloscope.
resultsEx vivo, VC had 100% TSP success with 1 attempt using 1 s (constant) and 300 ms (pulse) modes; EG demonstrated higher failure rates and less consistency at 30 W (78%) and 40 W (88%), requiring more RF applications, longer duration and energy output. EG crossings ex vivo and in vivo showed thermal damage on septa and device charring along the uninsulated distal end, unlike VC. Successful TSP with EG (1 s, 30-50 W) in vivo demonstrated wire charring and thrombus along the uninsulated wire length within the dilator.
conclusionManual protrusion of a wire with a long uninsulated length can be difficult to control in a clinical setting, thus, introducing variability, excess energy, and risk of thermal injury and thrombus formation. A purpose-built electrode and dedicated generator optimize RF delivery for efficient and safe TSP.