Bridging physics and practice: evaluating sensitivity, septal penetration, and detector dead time in terbium-161 gamma-camera imaging.
Article in EJNMMI physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Improving image quality in terbium-161 phantom imaging: Quantitative evaluation of DEW and TEW scatter correction methods.Journal of applied clinical medical physics · 2026Article
- QuantitativeEJNMMI physics · 2026Article
- Quantitative performance of a 3D CZT and conventional SPECT/CT for Tb-161 imaging.EJNMMI physics · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
INTRODUCTION/
aimTerbium-161 (
methodsMeasurements were conducted on a Siemens Symbia Intevo system using two collimators: low-energy high-resolution (LEHR) and medium-energy low-penetration (MELP). Two energy windows were evaluated: 75 keV ± 10% and 48 keV ± 20%. Planar sensitivity and penetration were assessed using a
resultsSensitivity was comparable for both collimators at 75 keV (LEHR: 15.7 cps/MBq, MELP: 18.5 cps/MBq) and increased at 48 keV (LEHR: 44.4 cps/MBq, MELP: 67.9 cps/MBq). Maximum penetration occurred at 75 keV with the LEHR collimator (7.5% at 10 cm). In acquired spectra, more than half of the detected counts (51.6%) appeared above the 75 keV window with LEHR, compared to only 12.2% with MELP. Dead-time analyses revealed non-linear detector responses at wide-spectrum count rates exceeding 93 kcps, corresponding to in-field activities of 1.4-2.0 GBq for LEHR and 1.7-2.2 GBq for MELP. The dead-time constant was determined to 0.42 µs for both detector heads, however, the maximum recorded count rate differed significantly (384 kcps vs. 546 kcps). The median and maximum wide-spectrum count rate for patients treated with [
conclusionsWhile high-quality