Evidence map›Paper›PMID 40857598›Full record

ArticleSICOT-J2025

A novel in vitro experimental design for biomechanical testing of patellofemoral joint kinetics and kinematics.

A Mounir Boudali, Jobe Shatrov, Koki Abe, Marcus Zavala, David Parker, William L Walter, Elizabeth Clarke

Abstract read
In one paragraph

Article in SICOT-J, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

A Mounir BoudaliMurray Maxwell Biomechanics Laboratory, Kolling Institute, Northern Sydney Local Health District, 10 Westbourne St, St Leonards, NSW 2065, Australia.ORCID 0000-0002-2737-581X
Jobe ShatrovLandmark Orthopaedics, Level 2, 500 Pacific Highway, St Leonards, NSW 2065, Australia.ORCID 0000-0003-3073-2416
Koki AbeMurray Maxwell Biomechanics Laboratory, Kolling Institute, Northern Sydney Local Health District, 10 Westbourne St, St Leonards, NSW 2065, Australia.ORCID 0000-0002-2894-624X
Marcus ZavalaUniversity of Miami, College of Engineering, Department of Mechanical and Aerospace Engineering, Advanced Materials Innovations Laboratory, 1600 NW 10th Ave., Miami, Fl 33136, USA.ORCID 0009-0006-1108-6876
David ParkerLandmark Orthopaedics, Level 2, 500 Pacific Highway, St Leonards, NSW 2065, Australia.ORCID 0000-0001-6930-8490
William L WalterMurray Maxwell Biomechanics Laboratory, Kolling Institute, Northern Sydney Local Health District, 10 Westbourne St, St Leonards, NSW 2065, Australia.ORCID 0000-0002-9413-0871
Elizabeth ClarkeMurray Maxwell Biomechanics Laboratory, Kolling Institute, Northern Sydney Local Health District, 10 Westbourne St, St Leonards, NSW 2065, Australia.ORCID 0000-0003-2500-5222

Funding

Number 9 trust fund Royal North Shore HospitalRamsay Health CareSmith and Nephew
6 · The paper itself

Abstract

introductionComplications arising from the patellofemoral joint (PFJ) represent the third most common cause for revision in total knee arthroplasty (TKA). Previous in vitro biomechanical studies have altered the native attachments of muscles controlling the PFJ. The purpose of this study was to design an in vitro biomechanical setup that would allow testing of both native and arthroplasty knee joints, specifically the PFJ, without disturbing the native attachments of the quadriceps and hamstrings muscles.

methodsAfter finalising a prototype, a pelvis-to-toe human cadaver specimen was tested. The simVITRO platform was used to simulate movement and control force trajectories. A motion capture system was used to capture the motion of the bones and to measure knee flexion angle and patellar movement with respect to the femur. The forces applied in the PFJ were measured using a custom patella sensor.

resultsDisplacement of the reflective cluster attached to the femur was measured during compression loading at different flexion angles, passive flexion and stairs descent trajectory. The femur showed less than 1 mm and 3 mm displacement with respect to the femur clamp in passive flexion and stairs descent. The most translation of 8.37 mm (<2% average femur length) was observed at 90° flexion which occurred at 483 N simulated compression force.

conclusionThis novel design provides a methodology for studying the biomechanics of the PFJ in vitro that preserves the soft tissues influencing the behaviour of the joint. This setup provides a biomechanics model that can be utilised to better understand and study the PFJ in vitro.

Indexed as

KneeKnee kinematicsKnee kineticsPatellofemoralPatellofemoral joint

Identifiers

PMID40857598
PMCPMC12380413

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.