Development and demonstration of end-to-end testing for intra-fraction motion-managed workflows.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Development and demonstration of end-to-end testing for intra-fraction motion-managed workflows.Medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundIntra-fraction motion management techniques, including beam gating and intra-fraction drift correction (IDC), have recently been introduced on the Unity MR-linac (Elekta AB, Stockholm, Sweden) to mitigate the dosimetric impact of motion during treatment. However, residual motion (e.g., within the gating window) still affects the delivered dose, causing deviations from the statically planned dose. Conventional end-to-end (E2E) testing does not incorporate such (known) motion, hampering evaluation of motion managed workflows. PURPOSE: This study develops and demonstrates novel methods that incorporate known motion before treatment delivery. Using such a reference dose distribution allows for E2E testing of intra-fraction motion-managed workflows.
methodsA novel approach was developed to assess the E2E accuracy for motion-managed delivery techniques by comparing the measured dose distribution to a reference dose distribution that incorporates the applied motion during the delivery. Two motion-included reference dose distributions were generated and evaluated: (1) A Priori Motion-Included (APriMI) dose distribution which uses the known (periodic) motion to estimate the influence of anatomical motion on the dose distribution, and incorporates this into a new dose distribution; and (2) the Posteriori Motion-Included (PostMI) dose distribution, which adds an external trigger to relate the beam-on/off time to the motion of the setup. This allows for evaluation of non-periodic motion, or a drift motion during IDC workflows. In addition to these, the conventionally used static treatment planning system (TPS) dose distribution was used as a reference dose distribution. Several scenarios were evaluated: static (no phantom motion), two unmanaged, and two motion-managed scenarios using the Comprehensive Motion Management (CMM) software (Elekta AB, Stockholm, Sweden) for gated and IDC workflows, with
resultsFirst, the static benchmark scenario was evaluated and showed high agreement between the measured dose distribution and all reference dose distributions (i.e., static, APriMI, and PostMI). For the motion-included scenarios, excellent agreement was observed between the measured and calculated dose distributions in both unmanaged and managed cases when using either APriMI or PostMI. The largest geometric shift in the motion included scenarios was 0.3 mm, comparable to the static scenario. Dosimetric accuracy, evaluated using a global gamma index (2%/2 mm), exceeded 95.5%. As expected, larger deviations occurred when the static dose distribution was used as a reference, with geometric shifts up to 9.0 mm and gamma pass rates as low as 17.8%.
conclusionsE2E testing of intra-fraction motion-managed workflows is possible using APriMI and PostMI dose distributions. Strong agreement was observed with these motion-included distributions, while larger deviations were seen with the static dose distribution. These findings highlight the need for reference dose files that account for actual motion in the measurement setup to assess E2E accuracy of motion-included workflows.