ArticleJournal of clinical monitoring and computing2026
Improved efficiency of the new Mirus™ system challenged by ventilatory settings-a bench study.
Article in Journal of clinical monitoring and computing, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Thermocycling: enhancing efficiency of anesthetic reflectors through cyclic heating and cooling-finding the optimal temperature in a bench study.Intensive care medicine experimental · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
The Mirus™ system (Technologie Institut Medizin GmbH, Germany) enables target-controlled delivery of volatile anaesthetics with open system ventilators. Its interface, positioned between Y-piece and the patient, injects anaesthetic vapor during inspiration, while adsorbing anaesthetic and resupplying it during the next inspiration. A newly introduced interface (Lisa-44) was evaluated under body temperature pressure saturated and normocapnic conditions using a test lung. Volume-controlled ventilation was applied with a 500 mL tidal volume, 10 bpm respiratory rate, and inspiratory flow (IF) times of 2.5 s and 0.5 s. Isoflurane (sevoflurane) target concentrations were stepwise increased to 1.5 MAC and then decreased. An external gas monitor recorded real-time anaesthetic concentrations. End-tidal concentrations (EC) and area under the curve (AUC) were analyzed for each breath. Accuracy and precision were assessed using Bland-Altman plots. Consumption was calculated and compared to historical controls. Mirus EC measurements correlated well with external gas monitor readings. ECs fluctuated around targets, with higher targets requiring more frequent injections and larger AUCs. Anaesthetic consumption was lower than historical controls and comparable to the Sedaconda ACD. At 0.3-0.5 MAC, hourly isoflurane consumption ranged from 1.0 to 1.7 mL (sevoflurane: 3.8-6.1 mL). However, very short IF times significantly increased consumption, reaching 43 mL sevoflurane at an EC of 2.6 vol%, with prolonged injections extending into the expiration phase. The new system demonstrates high accuracy, precision, and improved efficiency, suggesting reduced anaesthetic consumption in clinical use. However, high concentrations combined with very short IF times substantially increase consumption, indicating potential limitations.