ArticleBMC surgery2025
A 3D-printed screw mechanism as an alternative method to prevent wire migration in nonpalpable breast lesion localization.
Article in BMC surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundWire localization for nonpalpable breast lesions and pre-neoadjuvant chemotherapy (NAC) clips can lead to complications including wire migration, pneumothorax, and organ injury. This study evaluates the clinical efficacy and safety of a novel 3D-printed screw mechanism designed to prevent wire migration.
methodsThis retrospective descriptive study included 104 patients with nonpalpable breast lesions and pre-NAC clips who underwent wire localization between March 2023 and August 2024. A screw mechanism was designed and manufactured using 3D printing technology with polylactic acid (PLA) material (4 g per device). The device features a 21.7 mm base diameter and elliptical wire hole (2.5 × 4 mm). Wire stability and complications were assessed through imaging and clinical follow-up.
resultsThe median age was 47 years (range: 33-82) and median BMI was 24.7 (17.9-40.8). Lesions were primarily located in the upper outer quadrant (49.9%). Among 68 patients receiving NAC, 44.1% achieved pathological complete response (ypT0N0). The screw mechanism successfully prevented wire migration during patient transport from radiology to operating room in all 104 cases (0% migration rate, 95% CI: 0-3.6%). No complications including pneumothorax, organ injury, or skin reactions were observed. The device was easily removed by simple unscrewing on the operating table before surgical skin preparation. Median device application time was 2-4 h.
conclusionThe 3D-printed screw mechanism effectively prevents wire migration and demonstrates excellent safety profile. Its simple design and ease of application make it a practical solution for improving patient safety during breast lesion localization procedures.
trial registrationNot applicable. This retrospective study did not require trial registration.