An Investigation of the Dosimetric Uncertainties in the myOSLchip System: The Impact of Contributing Factors on Measurement Accuracy.

Lee C Goddard, Amar Basavatia, Patrik Brodin, Wolfgang A Tomé

Abstract read
In one paragraph

Article in Cureus, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Lee C GoddardRadiation Oncology, Montefiore Medical Center, Bronx, USA.
Amar BasavatiaRadiation Oncology, Montefiore Medical Center, Bronx, USA.
Patrik BrodinRadiation Oncology, Montefiore Medical Center, Bronx, USA.
Wolfgang A ToméRadiation Oncology, Montefiore Medical Center, Bronx, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The use of in vivo dosimetry (IVD) is integral to radiation oncology, particularly for treatments involving custom boluses or superficial electron lesions. In this report, the "myOSLchip" system, comprising beryllium oxide (BeO)-based optically stimulated luminescent chips and hand-held readers, was evaluated for clinical use. Each 4.7 × 0.5 mm chip, housed in plastic with an effective measurement point 1 mm below the surface, was calibrated under standard 6 megavoltage (MV) photon conditions using a 10 × 10 cm² field at 100 cm SSD and 1.5 cm depth in solid water. Calibration factors were assessed for energy dependence, dose linearity, time to readout, repeat measurements, and angular dependence. Results showed high measurement accuracy and linearity up to 10 Gy, with separate calibration factors needed per reader. The chips exhibited significant under-response at low energies (down to 16.82 keV), but minor variation in the megavoltage range for both photons and electrons. Repeat use showed an average signal loss of 1.2% after the first bleaching and 0.35% per subsequent cycle. A stable readout required at least four hours post-irradiation, with earlier readings (30-60 minutes) showing 2-3% error. No angular dependence was observed. With appropriate calibration and consideration of these factors, the myOSLchip system delivers dose measurements within the clinically accepted 3-5% accuracy range.

Indexed as

clinical validationenergy dependencein-vivo dosimetrymeasurement accuracymyosldosld

Identifiers

PMID40827195
PMCPMC12358160