Evidence map›Paper›PMID 40825559›Full record

ArticleBone & joint research2025

The influence of "kickstand" screws on the mechanical performance of a lateral distal femoral plate : a finite element analysis.

Alex Trompeter, Alexis Christen, Claus Gerber, Bernhard Hofstaetter, Felix Wipf, Jason Lowe

Abstract read
In one paragraph

Article in Bone & joint research, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Alex TrompeterCity St George's University of London, St George's University Hospital, London, UK.
Alexis ChristenStryker, Selzach, Switzerland.
Claus GerberStryker Trauma GmbH, Schönkirchen/Kiel, Germany.
Bernhard HofstaetterStryker Trauma GmbH, Schönkirchen/Kiel, Germany.
Felix WipfStryker, Selzach, Switzerland.
Jason LoweUniversity of Arizona, Banner - University Medicine, Alvernon Clinic, Tucson, Arizona, USA.

Funding

Stryker
6 · The paper itself

Abstract

Aims: Lateral locked plating of distal femoral fractures is widely reported, yet there remains a 9% to 19% incidence of mechanical failure. Obliquely directed "kickstand screws", from the metaphyseal portion of a plate toward far-sided articular subchondral bone, have been shown to improve construct stiffness. This study explores the impact of kickstand screws in a finite element analysis bone defect model, comparing plate and screw maximum stress and maximum locking screw forces either with or without the addition of kickstand screws. Methods: A finite element analysis model of a lateral based femoral plate and fracture gap simulation was created, with material and construct data parameters regarding bone material, implant, and composite model identified. The addition of the upper, lower, or both kickstand screws in an anatomical precontoured lateral distal femoral plate were selected as the variables compared against the absence of kickstand screws. Screw and plate principal stresses (MPa) and locking screw mechanism force (N) were measured. Results: The addition of the upper kickstand screw or both kickstand screws led to an approximate 40% reduction of stress in the metaphyseal hole closest to the fracture. The addition of the lower, upper, and both kickstand screws led to a 23%, 32%, and 34% reduction of maximum stress in the metaphyseal screws, respectively. The addition of the lower kickstand screw led to a 19% reduction, while the upper or both kickstand screws led to a 23% reduction of maximum force experienced by the locking mechanism. Conclusion: The addition of kickstand screws improves the mechanical performance of the construct, with reduced stresses experienced by the plate and metaphyseal screws. Furthermore, the maximum forces on the locking screw mechanism were shown to be significantly reduced, providing a protective effect to the polyaxial locking mechanism of the metaphyseal screw.

Identifiers

PMID40825559
PMCPMC12360867

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.