A novel Cherenkov radiation removal method for plastic scintillator detectors in a 0.35 T MR-Linac.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Pulse‑level characterization of low monitor unit deliveries on a modern linear accelerator using a plastic scintillation detector.Journal of applied clinical medical physics · 2026Article
- Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.Medical physics · 2026Article
- A novel Cherenkov radiation removal method for plastic scintillator detectors in a 0.35 T MR-Linac.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThis study evaluates methods for removing Cherenkov radiation (CR) from plastic scintillation detectors (PSDs), focusing on constraints specific to a 0.35 T MR-Linac system.
methodsFive CR-removal methods were examined: cross calibration, fiber alone, multiloop, collimator rotation, and couch rotation. The first three (cross calibration, fiber alone, and multiloop) were tested on a 0.35 T MR-Linac (ViewRay Inc., USA) using the BluePhysics PSD (Blue Physics LLC, USA). These methods do not require collimator or couch rotation. The remaining two methods (collimator rotation and couch rotation) were tested on a Varian TrueBeam (Varian Medical Systems, USA) for comparison. Measurements were performed under various setup configurations, and Cherenkov radiation extraction (CRE) values were calculated to determine each method's effectiveness.
resultsThe multiloop approach yielded a CRE of 0.7288, making it the most practical and robust for MR-Linac constraints because it requires neither collimator nor couch rotation. The cross calibration and fiber alone methods produced CRE values of 0.7318 and 0.7569, respectively. Collimator rotation gave 0.7255, comparable to multiloop. In contrast, couch rotation resulted in 0.7489 but exhibited more variability, suggesting lower reliability.
conclusionThe multiloop method emerged as the most practical and robust technique for CR removal in 0.35 T MR-Linac systems. Its simplicity and compatibility with MR-Linac design constraints make it a highly effective approach for CR removal in PSD-based radiotherapy applications.