Impact of magnetic fields on small field dosimetry for parallel MR-linacs.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Enhancement of electron beam conformity in MRI-guided radiotherapy with parallel magnetic fields: a Monte Carlo analysis.Radiological physics and technology · 2026Article
- Development and validation of a TOPAS Monte Carlo model for a 0.5 T bi-planar Linac-MR.Medical physics · 2026Article
- Impact of magnetic fields on small field dosimetry for parallel MR-linacs.Medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
1 author.
Funding
Abstract
backgroundThe use of MR-guided radiation therapy (MRgRT) is increasing with the introduction of commercially-available parallel MR-linac systems. Treatment fields used by these machines can be very small and highly modulated. Literature on the small field dosimetry of parallel MR-linacs and the impact of the magnetic field on dose perturbations is scarce. PURPOSE: To investigate the impact of magnetic fields on small field dosimetry of parallel MR-linacs using Monte Carlo (MC) methods.
methodsA TOPAS MC model of a 6 MV FFF beam was developed and validated against a commercial 0.5 T parallel MR-linac. The impact of varying magnetic field of up to 3.0 T on water phantom dosimetry such as percent depth dose (PDD), beam profiles, and output factors (OFs) was studied. A lung phantom with embedded spherical tumors of diameters 1-3 cm was employed to investigate the impact of parallel magnetic fields on lung lesions using an arc delivery with a 4
resultsNegligible differences were noted in the PDD, except for the build-up region, between the investigated parallel magnetic fields. Besides the tail region of the 1.5 and 3.0 T profiles, minimal differences were observed in the lateral beam profiles. Similarly, parallel magnetic fields of up to 1 T had a negligible impact on small field OFs. Compared to the 0 T magnetic field, the OF for field sizes of <2 × 2 cm
conclusionsThe impact of parallel magnetic fields of up to 1.0 T strength is minimal for in-water small field dosimetry. However, the increase in OF, compared to the 0 T case, was found to be significant for the 3.0 T magnetic field. For all field strengths compared to 0 T, significant dose differences were found around the periphery of the tumors inside a lung phantom that must be accounted for during treatment planning and dosimetry.