Biomechanical testing of three constructs for prosthetic laryngoplasty in horses demonstrates advantages of differing metallic implants in the arytenoid cartilage.
Article in Veterinary surgery : VS, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
Abstract
objectiveTo compare the biomechanical stability of a novel arytenoid implant (tie-bolt, TB) with a suture anchor (SA) and standard laryngoplasty using suture alone (control) in equine larynges. STUDY
designRandomized ex vivo controlled experimental study. SAMPLE POPULATION: Thirty equine cadaver larynges.
methodsLarynges were randomly assigned to TB, SA, or control groups (n = 10 each). The force of abduction was measured for each construct, then constructs underwent cyclic loading (5-50 N for 3000 cycles) to assess elongation, followed by load-to-failure testing to assess ultimate load, displacement, stiffness, and mode of failure.
resultsThe SA required less force to achieve 88% abduction (8.31 ± 2.46 N) than the TB (11.48 ± 3.6 N, p = .07) and controls (11.86 ± 2.99 N, p = .04). The TB had less construct elongation (2.62 ± 0.61 mm) than SA constructs (3.87 ± 1.25 mm, p = .01) and was stiffer (31.05 ± 3.38 N/mm) than both control (23.48 ± 2.06 N/mm, p < .0001) and SA constructs (18.53 ± 4.15 N/mm, p < .0001). The TB construct failed at higher ultimate load and displacement (367.25 ± 66.98 N and 19.11 ± 2.51 mm, respectively) than the control (238.14 ± 19.46 N and 14.23 ± 1.6 mm; p < .0001 for both) and SA constructs (196.55 ± 47.17 N and 16.23 ± 2.02 mm; p < .0001 and p = .01, respectively). The primary modes of failure were suture cut-through for control, implant pullout for SA, and arytenoid fracture for TB.
conclusionThe SA required the least force for abduction. The TB was a stiffer construct that resulted in higher failure loads. CLINICAL SIGNIFICANCE: The SA may reduce construct loading during abduction. The TB offers superior mechanical stability, potentially reducing long-term degradation from cyclic loading.