Article3D printing in medicine2025
Dimensional accuracy of 3D-printed surgical cutting guides after hospital sterilization: a comparative evaluation of ten MEX materials.
Article in 3D printing in medicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Clinical Implications of Sterilization Methods Applied to 3D-Printed Implant Surgical Guides: An In Vitro Study.International dental journal · 2026Article
- A Review of 3D-Printed Medical Devices for Cancer Radiation Therapy.Bioengineering (Basel, Switzerland) · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIntegrating 3D printing into orthopedic oncology enables the development of patient-specific cutting guides for specific anatomy. To preserve surgical precision, especially in tumor resections where the safety margins must balance minimization of recurrence with avoidance of excessive bone removal, it is critical to maintain the dimensional accuracy of these guides throughout all stages of fabrication, disinfection, cleaning, and sterilization.
methodsPersonalized cutting guides were 3D printed using ten filaments, and 3D scanned before and after sterilization. Two sterilization methods were used: autoclave and vaporized hydrogen peroxide. Dimensional deviations were assessed by comparing the reference STL model with the scanned models using metrics such as root mean square, standard deviation, Gaussian mean, and maximum error. Pearson correlation analysis was conducted to evaluate inter-sample variability and metric interdependence.
resultsPLA and PETG showed the best dimensional accuracy in the as-printed state with RMS values of 0.093 mm and 0.093 mm, respectively, and standard deviations below 0.092 mm. After hydrogen peroxide sterilization, PETG, PC, and PETG-CF kept a high accuracy, while PLA, PLA-HP, PA, and PA6-CF showed significant deformations. Autoclave sterilization determined severe deformation in most materials, with PC showing unexpectedly changes of the geometrical form, increasing in RMS error from 0.127 mm to 3.642 mm. In the as-printed state, maximum error remained below 0.29 mm for all materials, with PLA having the highest localized deviation (0.283 mm). After hydrogen peroxide sterilization, PETG, PC, and ABS maintained maximum error values lower than 0.27 mm, while PLA increased to 0.274 mm and PLA-HP to 0.268 mm. These values, although moderate, showed geometric changes that affect fit in anatomically constrained regions. Pearson correlation analysis showed that hydrogen peroxide sterilization altered the relationship between accuracy metrics of prints after manufacturing, weakening the correlation between RMS and Gaussian mean. This suggested increased unpredictability in deformation direction and highlighted less consistent deformation patterns.
conclusionsDisinfection and sterilization processes were highly material-dependent, as expected. PETG, PC, and PETG-CF were the most stable materials for the 3D-printed surgical guides when using cold plasma sterilization. Materials like PLA, PLA-HP, and PA require caution due to their instability. Designers should take into account the deformation directionality loss post-sterilization and integrate fit allowances into surgical guide geometry.