ArticleArtificial organs2025
Ghost Cells as a Two-Phase Blood Analog Fluid -Fluorescent Mechanical Hemolysis Detection.
Article in Artificial organs, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Ghost Cells as a Two-Phase Blood Analog Fluid-Optical Thrombus Growth Detection Using Particle Image Velocimetry.Artificial organs · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
backgroundThis study investigated fluorescent hemolysis detection as an optical method to detect local hemolysis in mechanical circulatory support systems, addressing the limitations of standard hemolysis tests and current simulation methods. Standard tests, per ASTM1841-19, quantify general hemolysis but do not localize it.
methodsWe employ a two-phase blood analog fluid composed of calcium-loaded ghost cells and phosphate-buffered saline. Ghost cells are hemoglobin-depleted red blood cells, allowing for optical measurements. A calcium-sensitive fluorescent indicator (Cal590 potassium salt, AAT Bioquest, Pleasanton, USA), activated by calcium released upon ghost cell hemolysis, enables fluorescent hemolysis detection. Hemolysis tests were conducted using porcine whole blood and the blood analog fluid, confirming that both undergo mechanical hemolysis in the Food and Drug Administration pump model.
resultsThe results revealed increased fluorescence intensity in response to hemolysis, with a quantitative fluorescence increase of 8.85/min at 3500 rpm and 2.5 L/min, indicating hemolysis, particularly at the rotor tip. Through image processing of fluorescence images, local hemolysis was visualized.
conclusionThis study is the first to use fluorescent hemolysis detection for local detection of mechanical hemolysis. Further refinement may enhance the design of mechanical circulatory support systems and bridge simulation limitations with experimental, localized hemolysis detection.