ArticleJournal of diabetes investigation2025
Study of insulin infusion in diabetic rats with an ultra-low-cost insulin pump for managing blood glucose levels.
Article in Journal of diabetes investigation, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
introductionThis study assesses the accuracy of an ultra-low-cost insulin pump for regulating blood glucose, an essential component of the artificial pancreas.
objectivesThe primary objective is to assess the accuracy of blood glucose regulation using an ultra-low-cost insulin pump compared with insulin injections.
methodsThe study was conducted using 14 male Wistar rats aged 14 weeks, which were induced with diabetes through a streptozotocin procedure. The control group (seven subjects) was treated with subcutaneous insulin injections, and the study group (seven subjects) was treated with the experimental insulin pump.
resultsOur findings reveal that the ultra-low-cost insulin pump performs comparably to insulin injections for glycemia regulation using standard clinical insulin treatment. The blood glucose records obtained by CGM indicate that trends in both methods correspond, with a glycemia median absolute relative difference of 6.1748 mg/dL and an interquartile range of 14.57 mg/dL. The response time to reach glucose levels between 70 and 180 mg/dL was around 55 min, testing different amounts of insulin units ranging from 0.09 to 2.06 U. Also, four obstruction cases were detected by the system in an average time of 3 s, showing compliance with the EN60601-2-24 standard.
conclusionsThe in vivo experiments with diabetic rats and ultra-low-cost insulin pumps demonstrate that these prototypes can be used for glucose regulation in clinical trials while complying with the conditions of the EN60601-2-24 standard. Future work should focus on conducting insulin pump tests in a closed loop and a cybersecurity study to prevent any possible external attack.