ArticlePractical laboratory medicine2025

Performance validation of Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.

Hongyu Zhang, Baixiu Wu, Liuhua Ke, Zheng Peng

Abstract read
In one paragraph

Article in Practical laboratory medicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Hongyu ZhangDepartment of Clinical Laboratory, The Third Affiliated Hospital of Guangxi University of Chinese Medicine, Liuzhou Traditional Chinese Medical Hospital, The Third Clinical Faculty of Guangxi University of Chinese Medicine, Liuzhou, Guangxi, China.
Baixiu WuDepartment of Nuclear Medicine Department, The Fourth Affiliated Hospital of Guangxi Medical University, Liuzhou Worker's Hospital, Liuzhou, Guangxi, China.
Liuhua KeDepartment of Clinical Laboratory, The Third Affiliated Hospital of Guangxi University of Chinese Medicine, Liuzhou Traditional Chinese Medical Hospital, The Third Clinical Faculty of Guangxi University of Chinese Medicine, Liuzhou, Guangxi, China.
Zheng PengDepartment of Clinical Laboratory, The Third Affiliated Hospital of Guangxi University of Chinese Medicine, Liuzhou Traditional Chinese Medical Hospital, The Third Clinical Faculty of Guangxi University of Chinese Medicine, Liuzhou, Guangxi, China.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Objective: To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests. Method: Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed. Results: Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from -0.4 % to 0.17 %, which met the requirement of less than 1 %. Conclusion: The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.

Indexed as

AlinityFT3FT4Performance validationTSHTT3TT4

Identifiers

PMID40677411