ArticleJournal of applied clinical medical physics2025
Commissioning and clinical implementation of low dose dual-field rotational TSET.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Commissioning and clinical implementation of low dose dual-field rotational TSET.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundTotal skin electron therapy (TSET) is a specialized technique for treating cutaneous T-cell lymphomas, such as mycosis fungoides. While traditional TSET regimens prescribe 30-36 Gy, low dose TSET at 12 Gy has gained attention due to its reduced toxicity. The dual-field rotational TSET technique, utilizing a standard linac, offers a practical solution for improved dose distribution. However, limited studies exist on its commissioning and clinical implementation. PURPOSE: This study aimed to commission and clinically implement a low dose dual-field rotational TSET technique. The primary objectives were to optimize beam geometry, characterize dosimetry, and ensure treatment uniformity while maintaining compliance with ACPSEM, EORTC, and AAPM guidance.
methodsA Varian TrueBeam linac equipped with a 6 MeV high-dose-rate total skin electron beam was commissioned for TSET. Dosimetric measurements included beam output calibration, relative dosimetry, and dose uniformity assessment using an ion chamber and Gafchromic film, with verification using an anthropomorphic phantom. A treatment dose calculation methodology was developed. Shielding, ozone generation, and quality assurance were assessed. In vivo dosimetry was performed for treatment validation.
resultsThe optimal dual-field beam geometry was determined to be ± 18° from the horizontal, achieving vertical uniformity within ± 2.9% and an overall treatment plane uniformity of ± 8.6%. Under rotation, the depth dose is delivered 100% at the surface, with bremsstrahlung contamination of 0.2 Gy. Custom eye shielding was developed, and ozone concentrations remained below the NCRP 0.1 parts per million (ppm) safety thresholds. In vivo dosimetry confirmed treatment uniformity within EORTC guidelines and identified three regions requiring dose boosts: the shoulders, palms, and inner thighs.
conclusionsThe commissioned dual-field rotational TSET technique provides a viable treatment option, achieving clinically acceptable dose distribution.