ArticleJournal of applied clinical medical physics2025

Commissioning and clinical implementation of low dose dual-field rotational TSET.

Broderick Ivan McCallum-Hee, Mounir Ibrahim, Godfrey Mukwada, Pejman Rowshanfarzad, Joshua Dass, Jude Dewitt, Robert Parin, Gregory Withey, Zaid Alkhatib

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Broderick Ivan McCallum-HeeDepartment of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.ORCID https://orcid.org/0000-0003-1595-9196
Mounir IbrahimDepartment of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.
Godfrey MukwadaDepartment of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.
Pejman RowshanfarzadSchool of Physics, Mathematics and Computing, The University of Western Australia, Crawley, Western Australia, Australia.ORCID https://orcid.org/0000-0001-8306-7742
Joshua DassDepartment of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.ORCID https://orcid.org/0000-0002-4726-9192
Jude DewittDepartment of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.
Robert ParinDepartment of Medical Technology & Physics, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.
Gregory WitheyDepartment of Medical Technology & Physics, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.
Zaid AlkhatibDepartment of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundTotal skin electron therapy (TSET) is a specialized technique for treating cutaneous T-cell lymphomas, such as mycosis fungoides. While traditional TSET regimens prescribe 30-36 Gy, low dose TSET at 12 Gy has gained attention due to its reduced toxicity. The dual-field rotational TSET technique, utilizing a standard linac, offers a practical solution for improved dose distribution. However, limited studies exist on its commissioning and clinical implementation. PURPOSE: This study aimed to commission and clinically implement a low dose dual-field rotational TSET technique. The primary objectives were to optimize beam geometry, characterize dosimetry, and ensure treatment uniformity while maintaining compliance with ACPSEM, EORTC, and AAPM guidance.

methodsA Varian TrueBeam linac equipped with a 6 MeV high-dose-rate total skin electron beam was commissioned for TSET. Dosimetric measurements included beam output calibration, relative dosimetry, and dose uniformity assessment using an ion chamber and Gafchromic film, with verification using an anthropomorphic phantom. A treatment dose calculation methodology was developed. Shielding, ozone generation, and quality assurance were assessed. In vivo dosimetry was performed for treatment validation.

resultsThe optimal dual-field beam geometry was determined to be ± 18° from the horizontal, achieving vertical uniformity within ± 2.9% and an overall treatment plane uniformity of ± 8.6%. Under rotation, the depth dose is delivered 100% at the surface, with bremsstrahlung contamination of 0.2 Gy. Custom eye shielding was developed, and ozone concentrations remained below the NCRP 0.1 parts per million (ppm) safety thresholds. In vivo dosimetry confirmed treatment uniformity within EORTC guidelines and identified three regions requiring dose boosts: the shoulders, palms, and inner thighs.

conclusionsThe commissioned dual-field rotational TSET technique provides a viable treatment option, achieving clinically acceptable dose distribution.

Indexed as

ElectronsLymphoma, T-Cell, CutaneousMycosis FungoidesParticle AcceleratorsPhantoms, ImagingRadiotherapy Planning, Computer-AssistedSkin NeoplasmsHumansOrgans at RiskQuality Assurance, Health CareRadiometryRadiotherapy DosageRadiotherapy, Intensity-Modulateddual field rotational TSETradiation oncologyradiotherapytotal skin electron irradiationtotal skin electron therapyTSEITSET

Identifiers

PMID40660928
PMCPMC12260257

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.