Geometric characterization of an electromagnetic surface tracking system in a radiation therapy environment.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Geometric characterization of an electromagnetic surface tracking system in a radiation therapy environment.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
13 authors.
Funding
Abstract
backgroundDespite advances in image-guided radiation therapy (IGRT), real-time, soft-tissue-based, volumetric motion monitoring remains unsolved. Integrated MRI+Linac systems are a solution, but are costly and complex. X-ray and optical photogrammetry-based systems have their limitations. Surrogate-based motion models, which use external signals to estimate internal motion, offer an alternative. We explore the feasibility of an electromagnetic (EM) fiducial-based device integrated with a surrogate-based motion model for real-time in-room volumetric motion monitoring. PURPOSE: To assess the feasibility of an EM-tracking system in the linac room, with an eventual goal of integrating it into an MRI-compatible system for real-time volumetric motion monitoring.
methodsWe empirically assessed the impact of gantry rotation and the radiation beam on EM-tracking accuracy using a sinusoidal motion trajectory (2 cm peak-to-peak, 5 s per cycle) programmed into a 2D motion platform. Four EM-tracking sensors were affixed to the platform, and their recorded trajectories were compared to the programmed motion under various conditions, including static and dynamic gantry positions, with and without radiation beams, and during CBCT acquisition.
resultsThe EM-tracking system faithfully reproduced the programmed sinusoidal motion during treatment beam (MV) and CBCT acquisition (kV + gantry rotation). With the beam off and static gantry and static motion platform at 0°, the average point-wise tracking difference was < 0.5 mm compared to gantry angles of 90°, 180°, and 270°. Similarly, with a moving platform, the sensors achieved a < 1 mm difference at the same angles. Additionally, the gantry's clockwise and anticlockwise rotations caused a < 0.5 mm difference on average at all angles during beam-off.
conclusionPreliminary results show the EM-tracking system operates with sub-millimeter accuracy in the linac room, with minimal effects from the radiation beam, gantry motion, or CBCT acquisition, supporting its feasibility for real-time volumetric motion monitoring during IGRT.