Trial reportArchives of orthopaedic and trauma surgery2025
Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses.
Trial report in Archives of orthopaedic and trauma surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 1 synthesis or guideline pooled it.
- A comparative meta-analysis of stress shielding between short-stem and conventional femoral stem in total hip arthroplasty.BMC musculoskeletal disorders · 2026Pooled it
- Bone-Sparing Hip Arthroplasty in Young and Active Patients: A Narrative Review.Journal of clinical medicine · 2026Review
- When the Tibia Hurts: Pain and Revision After Extendable Distal Femoral Endoprostheses in Young Patients.Indian journal of orthopaedics · 2026Article
- Early-term clinical and radiographic outcomes of a new short cementless stem for primary total hip arthroplasty in young and active patients.Journal of experimental orthopaedics · 2026Article
- Staged treatment of osteonecrosis of the femoral head induced by delayed treatment of polytrauma complicated by chronic acetabular fracture and hip dislocation: a case report.Frontiers in surgery · 2026Article
- Comparative analysis of the effects of blade plate retention versus removal on paediatric bone remodelling following proximal femoral osteotomy.Archives of orthopaedic and trauma surgery · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundShort-stemmed endoprostheses were developed to implement proximal load transmission and thus avoid stress-shielding in the proximal femur. Various prosthesis systems have been developed, which are discussed in the literature regarding stress shielding, clinical outcome, and long-term implant stability.
methodsIn this prospective randomized study, 52 patients (27 male, 25 female; average age 60.8 years) with conservatively unsuccessfully treated coxarthrosis were implanted with either a Nanos™ or Optimys™ short-stem prosthesis. Assessment included Gruen-zone based DEXA examinations immediately postoperatively and at one year to evaluate bone mineral density (BMD) and stress shielding, along with clinical outcomes using the Harris Hip Score (HHS). Radiographic measurements included offset (OFF), caput-collum-diaphyseal angle (CCD), leg length (LL), stem migration and inclination, and the occurrence of radiolucent lines (RL), assessed preoperatively, postoperatively, and at 12 months.
resultsDEXA showed differing stress-shielding profiles between stem types, favoring Optimys™ for BMD preservation. The Nanos™ group exhibited significantly greater BMD reduction in Gruen zones 1 (- 10.1%; p = 0.001), 4 (- 3.2%; p = 0.02), and 7 (- 21.3%; p = 0.001), whereas Optimys™ showed a significant decrease only in zone 7 (- 16.2%; p = 0.001). Although OFF, CCD, and LL changed significantly within groups postoperatively (p < 0.05), no statistically significant differences were found between the two stem designs in the final postoperative measurements (all p > 0.05). Stem migration remained clinically irrelevant in both groups. A statistically significant intra-group change was observed only in the Optimys™ group (Nanos™: 1.7 mm, p = 0.13; Optimys™: 2.5 mm, p = 0.01). Similarly, a small but statistically significant change in stem inclination was observed within both groups (Nanos™: 2.2°, p = 0.002; Optimys™: 1.5°, p = 0.01). Clinical improvement as measured by the Harris Hip Score (HHS) was excellent in both groups, with no significant differences between systems (Nanos™ pre/post: 52.0 / 98.0; Optimys™ pre/post: 51.6 / 97.0; both p < 0.001).
conclusionsWhen compared, the Optimys stem demonstrated reduced stress shielding through improved proximal load transmission, resulting in significantly better preservation of bone mineral density in the proximal femur. LEVEL OF EVIDENCE: Ib.
Indexed as