Investigation of the Response of Binary Multileaf Collimator Compensation to Target Setup Errors in the Radixact Synchrony System: A Phantom Study.

Masatoshi Miura, Koji Sasaki, Yasuo Shiota, Kazuyasu Inoue

Abstract read
In one paragraph

Article in Cureus, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Masatoshi MiuraDepartment of Medical Physics, Iwata City Hospital, Iwata, JPN.
Koji SasakiGraduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, JPN.
Yasuo ShiotaDepartment of Medical Physics, Iwata City Hospital, Iwata, JPN.
Kazuyasu InoueDepartment of Radiation Technology, Iwata City Hospital, Iwata, JPN.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The Radixact Synchrony system incorporates real-time target tracking into helical tomotherapy, dynamically correcting motion via jaw tracking and binary multileaf collimator (MLC) offset adjustments. Jaw tracking offers adjustments for longitudinal motion (IEC-Y), whereas MLC-based compensation for transverse motion (IEC-X and IEC-Z) is constrained by the 6.25-mm discrete leaf width, potentially affecting dosimetric accuracy. Furthermore, to formulate effective therapeutic approaches in motion-adaptive radiation therapy, understanding the MLC compensation threshold is crucial. This study aimed to assess the robustness of setup error correction and estimate the binary MLC compensation threshold in the Radixact Synchrony system through a phantom-based dosimetric analysis. Experiments were conducted using the Accuray Precision treatment planning system (version 3.3.1, Accuray, Sunnyvale, CA, USA) and the Radixact X9 (version 3.0.3, Accuray, Sunnyvale, CA, USA). A stereotactic body radiation therapy plan for lung cancer was delivered to a moving phantom using Radixact Synchrony. The phantom motion was simulated using the Delta4 Phantom+ and the Hexamotion stage. To evaluate the impact of setup errors on dosimetric accuracy, measurements were performed under longitudinal and transverse setup error conditions. Dosimetric accuracy was evaluated on the basis of the gamma pass ratio (GPR) and dose profile analysis, and the binary MLC compensation threshold was estimated. Along the longitudinal direction, jaw tracking effectively preserved dosimetric accuracy despite setup errors. Along the transverse directions, the GPR decreased as setup errors increased but improved when MLC compensation became active. Dose profile analysis revealed that MLC compensation was initiated at 4.3 mm. This study evaluated the impact of MLC compensation on dosimetric accuracy and estimated its threshold to be approximately 4.3 mm in the Radixact Synchrony system. Jaw tracking effectively corrects for longitudinal motion, whereas MLC-based compensation in the transverse direction is constrained by its discrete resolution and threshold activation. These findings offer valuable insights for refining treatment margins and enhancing motion-adaptive radiation therapy strategies. To enhance the accuracy and efficacy of Synchrony-based motion correction, studies focusing on clinical validation and parameter optimization are warranted.

Indexed as

dosimetric accuracymotion managementradixactreal-time adaptive radiotherapysynchronytomotherapy

Identifiers

PMID40621268
PMCPMC12227332

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.