Monitoring and optimization of POCT devices in a multi-specialty hospital in Poland: usage trends, quality assurance, and clinical impact (2017-2024).

Agnieszka Woźniak-Kosek, Lucyna Drążek

Abstract read
In one paragraph

Article in Acta biochimica Polonica, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Agnieszka Woźniak-KosekDepartment of Laboratory Diagnostics, Military Institute of Medicine - National Research Institute in Warsaw, Warsaw, Poland.
Lucyna DrążekDepartment of Laboratory Diagnostics, Laboratory of Medical Analytics in Legionowo, Military Institute of Medicine-National Research Institute in Warsaw, Warsaw, Poland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Rapid access to blood laboratory test results is crucial for diagnosing and treating patients in life-threatening conditions. Oxygenation status and acid-base balance are determined on arterial blood gasometry and are vital components of modern treatment algorithms for critically ill patients, similar to capillary blood glucose level measurement. The aim of this study is to present a modern method of reporting the work of critical parameter analyzers and glucometers in a multi-speciality hospital. The material for the analysis consisted of data obtained during the supervision of POCT devices in the clinics/departments/institutes of MIM-NRI from 2017 to 2024. Analyzing the use of glucometers in MIM-NRI from 2017 to 2024, it was noted that their usage significantly increased during the COVID-19 pandemic compared to the previous period. Currently, 112 devices are continuously used in both locations. All of them are subject to international and national quality testing, as well as daily internal calibration checking at their workplaces. There are currently 13 analyzers of critical parameter in the clinics and departments of MIM-NRI. All operate within the AQURE system, which allows for monitoring the correct operation of these devices and helps in quickly identifying analytical problems and reporting in real-time about the device's operational status. In the first half of 2024, 40,082 blood sample tests were performed on patients. Analysis of correctly performed gasometric tests conducted in clinics and laboratories shows similar values, ranging from 84.17% to 86.38%. In addition to external control, critical parameter devices must necessarily undergo daily internal laboratory control and calibration. The highest number of correctly performed calibrations, 9,239, was recorded in the Cardiac Surgery Clinic's Intensive Therapy Unit, accounting for 29% of all correctly performed calibrations for these devices. In the analyzed period of 2024, the rate of correctly performed quality checking's was 82.26%. The highest number of correct analyses, 636, was conducted in the Intensive Therapy Unit of the hospital in Legionowo, accounting for 10.32% of all controls performed. The demand for POCT tests in Polish hospitals is steadily growing due to their enormous potential and the time savings associated with performing tests directly at the patient's bedside. The greatest advantage of POCT is that by providing quick access to test results, diagnosis can be accelerated, and treatment initiated more rapidly. Additionally, POCT tests conducted in emergency rooms or hospital emergency departments can help to reduce the number of unnecessary hospitalizations or costly imaging tests. Having an appropriate POCT data management system in a multi-specialty hospital and ensuring IT communication is currently indispensable for proper patient care.

Indexed as

COVID-19Hospitals, SpecialPoint-of-Care SystemsPoint-of-Care TestingQuality Assurance, Health CareBlood Gas AnalysisBlood GlucoseHumansPolandSARS-CoV-2Blood Glucosecritical parameter analyzersglucometerslaboratory managementpoint-of-care testing (POCT)quality management

Identifiers

PMID40567349
PMCPMC12187598

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.