ArticlePloS one2025
Effect of autoclaving process on trueness and qualitative marginal fit of intraoral scan body manufactured from two materials: An in vitro study.
Article in PloS one, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Sterilization Effect on Axes and Distances Between Three Implants in Curved Lines When Using Two Types of Two-Piece Intra-Oral Scan Bodies: An In Vitro Study.Dentistry journal · 2026Article
- Dimensional Stability of Polymer and Titanium Implant Scan Bodies After Repeated Steam Sterilization: A High-Resolution 3D Metrological In Vitro Study.Journal of functional biomaterials · 2026Article
- PEEK Intraoral Scan Bodies-A Scoping Review.Dentistry journal · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
statement of the problemIt is not clear the effect of multiple autoclaving processes on the accuracy of scan bodies. Autoclaving dental disposables enables reuse, reducing the environmental impact of raw materials and waste. PURPOSE: This in vitro laboratory study evaluated the effect of the autoclaving process on the intraoral scan body (SB) manufactured from different materials (Titanium and PEEK), positioned at the implant and abutment levels.
methodsTwo models, each with 10 implants, were created to evaluate scan body trueness made of two materials (Titanium and PEEK), and implant junction positioned at two levels (implant and abutment levels). Analyses were conducted at five distinct time points: T0 (control group with new SB), T1 (after one autoclaving cycle), T2 (10 cycles), T3 (50 cycles), and T4 (100 cycles). Forty SBs were divided into four groups (n = 10), fixed on the models according to the initial positioning, using the recommended torque, and scanned 10 times using an intraoral Sccaner (InEOS X5 scanner, Dentsply Sirona). The surface deviation was evaluated by .STL mesh overlay using an image software (Geomagic Control X software). Microscopic analyses of marginal adaptation were performed simultaneously with the digital analysis using an optical microscope (Mitutoyo™-500 Optical Microscope) with 40x magnification at four places (mesial, distal, buccal, and lingual). Statistical analysis was performed using statistical software (SPSS 29.0 software), with α = 0,05. Data were subjected to the Shapiro-Wilk normality test. To assess spatial variation after the cycles, Repeated Measures ANOVA with Bonferroni correction was used. For microscopic qualitative analysis, percentile was used to check the prevalence of clinically adapted and misfitting faces.
resultsStatistically significant differences in mesh deformation were detected, with PEEK SB showing greater deformation compared to Titanium, especially at the abutment level, after 100 cycles (P < 0.01). Microscopic analysis revealed that 100% of the faces were classified as "Clinically Adapted" by the three calibrated evaluators (KAPPA > 0.08).
conclusionsTitanium scan bodies showed less surface deviation than PEEK but both remained below 50 µm. Internal-junction scan bodies were more affected than abutment-level ones. All scan bodies stayed within acceptable limits and viable after 100 autoclaving cycles.