ArticleIEEE transactions on bio-medical engineering2026
A Synergistic Patient-Specific Approach for Enhanced Spinal Fixation Using a Novel Flexible Pedicle Screw and a Complementary Steerable Drilling Robotic System.
Article in IEEE transactions on bio-medical engineering, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
objectiveCurrent spinal fixation (SF) techniques face screw loosening and pullout challenges in osteoporotic patients. This can be attributed to conventional rigid pedicle screws (RPS) being forced to fixate along a constrained linear trajectory into low bone mineral density (BMD) areas of the vertebral body. This study proposes a synergistic patient-specific approach that integrates a steerable drilling robotic system with a novel Flexible Pedicle Screw (FPS) to enhance SF procedures by enabling curved screw fixation.
methodsA patient-specific framework and synergistic design flowchart were developed to guide the synergistic design of the previously proposed Concentric Tube-Steerable Drilling Robot (CT-SDR) and the FPS. After, the novel FPS is designed based on critical design features and its design is validated using Finite Element Analysis (FEA). The FPS is then fabricated via Direct Metal Laser Sintering (DMLS). The FPS's morphability and self-tapping capability were experimentally assessed in Sawbones phantoms drilled by the CT-SDR system.
resultsThe FPS successfully morphed to fixate in curvilinear paths, demonstrating effective morphability and self-tapping in simulated bone.
conclusionBy enabling a flexible, patient-specific approach to pedicle screw fixation, the FPS and CT-SDR system address key limitations of current SF procedures. This method enhances screw anchorage and fixation strength in osteoporotic vertebrae. SIGNIFICANCE: This work presents a transformative approach to SF, with potential clinical applications in improving surgical outcomes for osteoporotic patients. The integration of robotic-assisted drilling and flexible implants could significantly reduce fixation failure rates, advancing orthopedic and spinal surgical practices.