ArticleRadiation oncology (London, England)2025
Long-term performance evaluation of a 1.5T MR-Linac using statistical process control techniques.
Article in Radiation oncology (London, England), 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- The impact of sex on setup accuracy in radiotherapy for rectal cancer: a real-World observational study.Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] · 2026Observational
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
Abstract
backgroundThe integration of magnetic resonance imaging with linear accelerators (Linacs) enhances adaptive radiotherapy by providing real-time imaging for improved treatment precision. However, the long-term performance of MR-Linac systems, particularly in clinical settings, remains insufficiently studied. Traditional quality assurance (QA) methods, relying on binary pass/fail criteria, may overlook critical system variations. This study applies statistical process control (SPC) techniques to evaluate the long-term performance of a 1.5T MR-Linac, focusing on optimization in beam quality, MR-to-MV alignment, MR imaging, and geometric distortion.
methodsA dual-phase SPC framework was applied to 1 year of daily and weekly QA data from an Elekta Unity MR-Linac. Phase I established performance benchmarks, while Phase II monitored deviations online. Evaluated parameters included beam output, symmetry, MR-to-MV alignment, signal-to-noise ratio (SNR), spatial linearity, slice profile, and geometric distortion across spherical volumes (DSVs). Stability and variability were quantified using control charts and process performance indices (Ppk).
resultsBeam quality was stable overall (Ppk ≥ 1.33), though output dose and transverse symmetry showed increased variability in Phase II, with dose Ppk declining from 3.13 to 1.33. MR-to-MV alignment was consistent, but Phi rotational and Z translational offsets showed variability after system upgrades. Imaging metrics, including SNR and spatial linearity, achieved A + performance (Ppk ≥ 1.67) in Phase II, while vertical spatial resolution was lower (Ppk 1.04-1.10). Geometric distortion was well-controlled, though larger DSVs (≥ 500 mm) showed increased AP-axis distortion (2.44 mm) compared to RL (1.37 mm) and FH (0.93 mm).
conclusionsSPC techniques dynamically identified stable parameters and areas for improvement. Key recommendations include enhanced alignment protocols for beam quality and MR-to-MV offsets, as well as targeted strategies to address geometric distortion in larger volumes and along the AP axis.