Observational studyActa diabetologica2025
Use of the temporary glycemic target feature in the medtronic MiniMed™ 780G advanced hybrid closed loop system: a real-world evidence study.
Observational study in Acta diabetologica, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Review
- Safety and glycaemic control during exercise with the MiniMed™ 780G: A narrative review of evidence from lab to real world.Diabetes, obesity & metabolism · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
aimThe MiniMed™ 780G system utilizes an Advanced Hybrid Closed-Loop algorithm, which is optimized to perform best at a glucose target of 100 mg/dL. The system also includes a temporary target feature (TT) of 150 mg/dL that can be set when higher glucose utilization is anticipated (e.g., exercise/strenuous work).
methodsReal-world data from 1002 randomly selected users of the MiniMed™ 780G and MiniMed™ 670G systems were used to investigate the TT functionality (n = 501 MiniMed™ 780G users, n = 501 MiniMed™ 670G users). Continuous glucose monitoring (CGM) metrics were calculated for 4-h intervals (4 h prior to commencing TT use and 0-4, 4-8, 8-12, 12-16, 16-20 and 20-24 h after TT initiation).
resultsCGM results showed that, with both systems, the time-in-range (TIR) during TT periods mostly equalled or exceeded TIR in corresponding non-TT periods. In the first 4 h after TT initiation, the MiniMed™ 780G system demonstrated a TIR and time in tight range (TITR) of 71.6% and 46.4%, respectively, [a 1.6 (p = 0.11) and 2.6 (p = 0.009) percentage point benefit in TIR and TITR, respectively, versus the 670G system]. Mean time-below-range remained below target in both periods.
conclusionsThese findings confirm that the TT setting performs as intended, allowing users to maintain TIR during strenuous activity without compromising safety.