ArticleJournal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing2026
Optimal power settings have major impact on procedural efficiency in pulmonary vein isolation guided by robotic magnetic navigation.
Article in Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Clinical value of robot-assisted catheter navigation for cardiac electrophysiology ablation: current evidence and future directions.Journal of robotic surgery · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionEarly reports on pulmonary vein isolation (PVI) for atrial fibrillation (AF) guided by robotic magnetic navigation (RMN) show comparable results to manual radiofrequency (RF) or cryoballoon ablation; however, lengthy procedures were reported. This could be due to suboptimal lesion formation caused by a lack of user experience and defined best practices. Operators performed RMN-guided RF ablation with lower power settings, presumably leading to longer procedures. In this study, we aimed to re-evaluate safety and efficiency of RMN-guided PVI for AF based on delivered radiofrequency power.
methodsPatients undergoing RMN-guided ablation were screened, and consecutive patients with AF undergoing PVI-only between 2008 and 2023 were retrospectively enrolled. Patients were grouped by the power settings utilized during the PVI procedure (group 1-25-30W, group 2-30-35W, group 3-35-40W, group 4-40-45W, group 5-45-50W, and group 6-50-55W). We collected and analyzed demographic data: age, sex, and AF type; safety data: intra- and post-procedural complications; procedural data: procedure duration, fluoroscopy time, RF ablation time, RF application number, and success rate; and follow-up data: AF recurrence and number of redo procedures.
resultsFrom the total number of 3398 screened patients, 238 patients met the inclusion criterion of undergoing PVI-only procedure (mean age 60.4 ± 9.9 years, 63.8% male). Throughout the 15 years only five patients had major (2.2%) and 15 patients had minor complications (6.6%), without differences between the patient groups (p = 0.40 and p = 0.63). The mean procedure duration was progressively decreased with the use of higher RF power (273.9 ± 97.0, 179.8 ± 104.0, 134.9 ± 55.3, 134.0 ± 39.5, 118.1 ± 41.3, and 110.9 ± 39.0 min, respectively; p < 0.001). Median fluoroscopy time was 19.5 min (IQR 13.0-35.5), progressively decreasing within the power groups (58.2 ± 20.5, 40.5 ± 26.2, 15.9 ± 6.6, 17.8 ± 8.1, 17.4 ± 7.5, and 19.8 ± 9.3 min; p < 0.001). We found differences between the power groups in RF application number (p < 0.001) and RF application duration (p = 0.003). Successful PVI was achieved in 238 patients (100.0%). Twenty-one patients with paroxysmal AF (17.1%) and 31 patients with persistent AF (40.7%) had documented recurrence during the 12-month follow-up. We found no differences in AF recurrence between the patient groups (p = 0.18 and 0.66).
conclusionsRMN-guided PVI-only for AF is safe and feasible. In contrast to early reports, procedure times and fluoroscopy use gradually decreased during the years, when increasing RF power was applied. Higher power settings during robotically-guided PVI did not compromise the safety of the procedures.