Mitigation of image distortion during mechanical testing within a dynamic stereo x-ray system.

David N Paglia, John M Chomack, David V Herlihy, Charlene Wetterstrand, Yazan Kadkoy, Roman Duchnycz, Patrick Kelly, J Patrick O'Connor, Susan E D'Andrea, Jason T Maikos

Registry-linked trialAbstract read
In one paragraph

Article in Frontiers in bioengineering and biotechnology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05287646 (Quantifying Bone and Skin Movement in the Residual Limb-Socket Interface of Individuals With Transtibial Amputation Using Dynamic Stereo X-Ray), which is not on this map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT05287646 narecruitingnot on this map

Quantifying Bone and Skin Movement in the Residual Limb-Socket Interface of Individuals With Transtibial Amputation Using Dynamic Stereo X-Ray

TypeinterventionalSponsorVA Office of Research and DevelopmentRan2024 to 2027Enrolled21ConditionsAmputation, Lower LimbArmsSuction Suspension First, Elevated Vacuum Suspension First
3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

David N PagliaDepartment of Orthopaedics, Rutgers-New Jersey Medical School, Newark, NJ, United States.
John M ChomackProsthetics and Sensory Aids Service, Veterans Affairs New York Harbor Healthcare System, New York, NY, United States.
David V HerlihyNarrows Institute for Biomedical Research and Education, Inc., Brooklyn, NY, United States.
Charlene WetterstrandDepartment of Orthopaedics, Rutgers-New Jersey Medical School, Newark, NJ, United States.
Yazan KadkoyDepartment of Orthopaedics, Rutgers-New Jersey Medical School, Newark, NJ, United States.
Roman DuchnyczDepartment of Orthopaedics, Rutgers-New Jersey Medical School, Newark, NJ, United States.
Patrick KellyDepartment of Orthopaedics, Rutgers-New Jersey Medical School, Newark, NJ, United States.
J Patrick O'ConnorDepartment of Orthopaedics, Rutgers-New Jersey Medical School, Newark, NJ, United States.
Susan E D'AndreaDepartment of Kinesiology, University of Rhode Island, Kingston, RI, United States.
Jason T MaikosProsthetics and Sensory Aids Service, Veterans Affairs New York Harbor Healthcare System, New York, NY, United States.

Funding

RRD VA I01 RX003700
6 · The paper itself

Abstract

Background: Dynamic stereo x-ray (DSX) permits Methods: A MuMETAL-lined box contoured to the test frame was developed to divert the magnetic field from the DSX test space. To assess the design, a radiopaque object was placed in the eMTS with shielding and within the DSX capture volume at either 65 or 103 cm from the image intensifiers (IIs) while the speed of the eMTS actuator was systematically increased from 0.1 to 10 mm/s during image collection. Root mean square error (RSME) was calculated over 1,000 frames for each test condition. Results: Results indicated a proportional change in RSME with increasing distance and decreasing speed. At 65 cm, higher actuator speeds (10 mm/s) produced the largest RSME (0.11 mm), significantly higher than the control test. At 103 cm, RSME was below 0.05 mm for all speeds. Conclusion: While closer distance to the IIs and higher actuator speeds produced larger RSME, results indicated that RSME for all experimental conditions fell below the established RSME associated with DSX marker tracking. The MuMETAL-lined box therefore mitigated DSX image distortion caused by the eMTS regardless of distance to the IIs and actuator speed. Clinical Trial Registration: clinicaltrials.gov, identifier NCT05287646.

Indexed as

biplanar fluoroscopydynamic stereo x-raymagnetic fieldmagnetic field shieldingmechanical testingskin strain

Identifiers

PMID40458257
PMCPMC12127421

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.