Evidence map›Paper›PMID 40444684›Full record

ArticleASAIO journal (American Society for Artificial Internal Organs : 1992)2026

Enhancing Left Ventricular Assist Device Usability: A Comparative Simulation Study of CorWave and HeartMate 3 Peripherals.

Gregor Widhalm, Theodor Abart, Katharina Ebenberger, Angelika Berger, Roxana Moayedifar, Daniel Zimpfer, Julia Riebandt, Michael Wagner, Thomas Schlöglhofer

Abstract readComparative Study
In one paragraph

Article in ASAIO journal (American Society for Artificial Internal Organs : 1992), 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Gregor WidhalmFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.ORCID 0000-0003-4762-5035
Theodor AbartFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.ORCID 0000-0002-1871-1834
Katharina EbenbergerFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.
Angelika BergerDivision of Neonatology, Pediatric Intensive Care and Neuropediatrics, Department of Pediatrics, Comprehensive Center for Pediatrics, Medical University of Vienna, Vienna, Austria.ORCID 0000-0001-8775-2405
Roxana MoayedifarFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.ORCID 0009-0001-3400-3248
Daniel ZimpferFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.ORCID 0000-0002-2185-4895
Julia RiebandtFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.ORCID 0000-0002-8601-9989
Michael WagnerDivision of Neonatology, Pediatric Intensive Care and Neuropediatrics, Department of Pediatrics, Comprehensive Center for Pediatrics, Medical University of Vienna, Vienna, Austria.ORCID 0000-0003-0982-2310
Thomas SchlöglhoferFrom the Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.ORCID 0000-0003-4354-4860

Funding

CorWave SA
6 · The paper itself

Abstract

Left ventricular assist devices (LVADs) improve outcomes but often compromise quality of life (QoL) due to usability issues with wearables. This study compared the usability of CorWave LVAD (CW) peripherals prototypes to HeartMate 3 LVAD (HM3) peripherals through a cross-sectional, simulation-based approach involving LVAD-naive laypeople, and heart transplantation (HTX) patients post-LVAD support. Simulations encompassed six tasks, measuring initial success rates, duration to success, pump-off time, and a post-scenario survey. Forty-six untrained participants (16 CW vs. 30 HM3; 12.5% vs. 20% female, p = 0.69; 56.2% vs. 50.0% HTX patients, p = 0.76) completed 276 scenarios. The CW cohort demonstrated superior initial success rates (92.7% vs. 80.6%, p = 0.008). Battery exchanges (in normal and dim light, within carry bag) were completed twice as quickly for CW users ( p ≤ 0.001). Although controller exchange success rates were comparable, the duration to success and pump-off times were doubled for the HM3 cohort ( p ≤ 0.005). During the "connection to AC power" scenario, HM3 subjects experienced higher complexity, as the CW cohort achieved 5× lower duration to success and 3× higher initial success rates ( p ≤ 0.001). Survey responses favored CW cable lengths (92.9% vs. 69.0%, p = 0.001). This study highlights the advantages of CW's user-centered design, which may enhance QoL and safety for future LVAD patients.

Indexed as

Heart-Assist DevicesHeart FailureAdultCross-Sectional StudiesFemaleHeart TransplantationHumansMaleMiddle AgedQuality of Lifehuman factors engineeringleft ventricular assist deviceusabilityuser-centered designwearables

Identifiers

PMID40444684
PMCPMC13021134

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked