Localization accuracy of 6-second CBCT for lung IGRT with various breathing patterns.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Development of CBCT-Guided Online Adaptive Radiotherapy.Cancers · 2026Review
- Evaluation of tumor localization accuracy on fast ring-gantry cone-beam computed tomography using patient-specific breathing curves and a dynamic anthropomorphic thorax phantom.Physics and imaging in radiation oncology · 2026Article
- Structure visibility and online adaptation suitability of a new high-performance ring-gantry cone-beam computed tomography imaging system in thoracic radiotherapy.Physics and imaging in radiation oncology · 2026Article
- Impact of tumor motion on target delineation and dose calculation accuracy using rapid-acquisition HyperSight CBCT in online adaptive radiotherapy.Journal of applied clinical medical physics · 2026Article
- Localization accuracy of 6-second CBCT for lung IGRT with various breathing patterns.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe 6-second cone beam computed tomography (CBCT) acquisition of the Ethos HyperSight (Varian Medical Systems, Inc. Palo Alto, CA, USA) on-board imaging system offers benefits, but could be too fast to accurately capture an average target position in a free-breathing lung cancer patient. This study aimed to ascertain whether a 6-second acquisition is appropriate for regularly breathing patients with varying respiration periods. Additionally, breathing patterns that could lead to clinically impactful inaccuracies for image-guided radiation therapy (IGRT) shifts were investigated.
methodsNine regular (sinusoidal) breathing patterns with different respiration rates (8-20 breaths-per-minute) and amplitudes, along with five irregular breathing patterns including a gasp/cough scenario, were tested using a respiratory motion phantom with a 3 cm diameter spherical target. Once the phantom was aligned using the external chassis to remove any default shifts, the 6-second CBCTs with Acuros reconstruction were acquired. Rigid registrations were performed using the 4DCT simulation average/untagged reconstructed image as reference to align to target (simulating clinical lung IGRT setup). Sixty-second CBCTs were also tested to verify the average offsets with conventional practice. The IGRT shifts were compared to 5 mm, a PTV margin typically used for lung tumors.
resultsMinimal (<1.0 mm) shifts were observed for all regular breathing patterns with both 6-second and 60-second CBCT acquisitions. For moderately irregular breathing patterns and the gasp/cough scenario, all shifts were less than 5 mm. Shifts larger than 5 mm were observed in highly irregular breathing patterns, with both 6-second (14%) and 60-second (24%) CBCT acquisitions. Statistical tests showed no significant differences (p > 0.05) between the sizes of shifts made with the two CBCT durations.
conclusionThe 6-second CBCT can effectively and reliably localize a mobile target for regular and moderately-irregular breathers. Cautions should be used for highly irregular breathers, regardless of the duration of CBCT acquisition.