ArticleEJNMMI physics2025
Dual-ended readout TOF-DOI PET detectors based on 3.2 mm and 1.6 mm pitch LYSO arrays.
Article in EJNMMI physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- High-Speed Detection of X-ray Pulses Using a Digital Counter Coupled with Perovskite Composite Scintillators.ACS applied materials & interfaces · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundThe image quality of positron emission tomography (PET) can be significantly enhanced by using time-of-flight (TOF) and depth-of-interaction (DOI) information. PET detectors are pivotal in determining the TOF and DOI capabilities of PET scanners.
methodsThis study developed and evaluated TOF-DOI PET detectors based on the dual-ended readout method and lutetium-yttrium oxyorthosilicate (LYSO) arrays with two different pitches and reflector configurations. Specifically, the performance of detectors based on three types of LYSO arrays with 20 mm thickness, 8 × 8 arrays with a 3.2 mm pitch, 16 × 16 arrays with a 1.6 mm pitch and normal reflectors, and 16 × 16 arrays with a 1.6 mm pitch and partial short reflectors, were assessed. Hamamatsu S14161-3050-08 silicon photomultiplier arrays were used as the photodetectors, and PETsys TOFPET2 was used as the readout electronics.
resultsThe flood histograms showed that all crystals in the three types of LYSO arrays were clearly resolved. The detectors based on the 8 × 8 LYSO arrays provided a coincidence timing resolution (CTR) of 207 ± 5 ps and a DOI resolution of 3.9 ± 0.6 mm. The detectors based on the 16 × 16 LYSO arrays with normal reflectors provided a CTR of 218 ± 7 ps and a DOI resolution of 2.6 ± 0.2 mm. In comparison, the detector based on the 16 × 16 LYSO arrays with partial short reflectors provided a CTR of 228 ± 11 ps and a DOI resolution of 2.9 ± 0.3 mm, and superior crystal resolvability compared to the detectors based on the 16 × 16 LYSO arrays with normal reflectors.
conclusionThese detectors are promising candidates for developing whole-body and brain PET scanners, offering effective sensitivity and uniform spatial resolution improvements across the field-of-view.