ArticleJournal of orthopaedic surgery and research2025
Biomechanical comparison of multiple zero-profile systems in anterior cervical discectomy and fusion: a finite element analysis.
Article in Journal of orthopaedic surgery and research, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Biomechanical evaluation of a novel cervical interfacet self-locking cage versus ACDF: a goat cadaveric study combined with human CT-based finite element analysis.BMC musculoskeletal disorders · 2026Article
- The design and biomechanical study of the 3D-printed implanted cervical vertebral cage.Journal of orthopaedic surgery and research · 2026Article
- Is zero-profile fixation safe in consecutive two-level anterior cervical discectomy and fusion with partial uncinate resection? A finite element analysis.BMC musculoskeletal disorders · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
backgroundAnterior cervical discectomy and fusion (ACDF) with zero-profile (ZP) implant is commonly used for cervical degenerative diseases, but subsidence remains a concern, particularly in osteoporosis. The two-screw ZP (TSZP), four-screw ZP (FSZP), and ROI-C implants are frequently applied, yet the biomechanical performance across varying bone qualities remains unclear.
methodsA finite element (FE) model of the cervical spine (C3-C7) was constructed with TSZP, FSZP, and ROI-C implants at C4/C5 to simulate normal and osteoporotic conditions. A 73.6 N load and 1 Nm torque were applied at C3 to simulate flexion, extension, lateral bending, and axial rotation, followed by biomechanical analysis.
resultsThe FSZP implant exhibited the smallest ranges of motion, followed by ROI-C, with the largest in TSZP. ROI-C showed the lowest peak implant system stresses, while TSZP had the highest on the anchoring device and FSZP on the cage. The TSZP implant had the highest cortical endplate stresses, whereas FSZP had the lowest in normal and ROI-C in osteoporosis. No significant differences were observed in adjacent intervertebral disc pressures. All parameters increased in osteoporosis, except cortical endplate stresses.
conclusionThe FSZP implant provided superior stability, while ROI-C exhibited a lower risk of implant-related complications. The TSZP implant was more prone to subsidence, which may be mitigated by optimizing stress distribution and enhancing damage prevention. Biomechanical performance was poorer under osteoporotic conditions, highlighting the need for careful surgical planning.