ArticleARYA atherosclerosis2025
Evaluation of novel ultrathin, biodegradable polymer tetriflex (sirolimus-eluting stent) optimization using intravascular ultrasound (IVUS) in short coronary lesion (≤ 20mm) vs. long coronary lesion (≥ 20mm): Tetriflex IVUS study.
Article in ARYA atherosclerosis, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Stent Undersizing and Malapposition: Mechanical Causes of Acute and Late Stent Failure.Reviews in cardiovascular medicine · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIntravascular ultrasound (IVUS)-guided percutaneous coronary intervention (PCI) is useful for stent optimization. Outcomes of the ultrathin Supralimus Tetriflex stent (Sahajanand Medical Technologies Pvt. Ltd., India) using IVUS were evaluated among patients with short (≤ 20 mm) versus long lesions (≥ 20 mm).
methodsA total of 207 patients underwent PCI, and IVUS was performed post-deployment. The primary outcome was optimal stent deployment, defined as (a) mean surface area (MSA) >5.0 mm²; (b) plaque burden <50%; (c) complete apposition; and (d) no edge dissection. Secondary outcomes were target lesion failure (TLF)-a composite of cardiac death, target vessel myocardial infarction (TVMI), and target lesion revascularization (TLR)-stent thrombosis, and major adverse cardiovascular events (MACE; a composite of death, MI, stent thrombosis, and repeat revascularization).
resultsSuboptimal deployment was significantly more frequent among patients with longer lesions (30.1% vs. 23.3%; p=0.03) due to higher rates of malapposition (17.3% vs. 10.6%) and MSA <5 mm² (9.6% vs. 7.7%). Following post-dilatation, suboptimal deployment was observed in 7.6% and 5.8% of patients, respectively. Residual plaque burden was 4.5% and 5.7%, respectively. The MSA in both groups was 6.3 mm² and 6.5 mm². Minimum and mean stent expansions were 82.1% versus 81.7% and 106.3% versus 109.8%, respectively, with no significant differences. TLF and stent thrombosis were observed in 0.9% versus 0.9% and 2.9% versus 3.8% of patients, respectively, with no significant differences. However, MACE was significantly higher (10.5% vs. 8.7%; p=0.05) among patients with longer lesions.
conclusionSupralimus Tetriflex stent has very good optimal deployment based on angiogram and becomes better with IVUS imaging, making it safe among long lesions (≥ 20mm).