ArticleNeonatology2025
Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates.
Article in Neonatology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- A Wireless Neonatal Intensive Care Unit: Fiction or Closer Reality?Neonatology · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
Abstract
introductionAccurate and continuous blood pressure (BP) monitoring in neonates is crucial in the intensive care unit. Invasive arterial lines (IALs), oscillometric cuffs, and current noninvasive continuous BP monitoring devices have significant limitations. The Boppli® device is a novel, continuous, noninvasive BP device that requires no calibration, designed for neonates.
methodsThis prospective, multicenter study evaluated the performance, usability, and safety of the Boppli device in neonates <5 kg. We compared mean arterial pressure (MAP), systolic blood pressure (SBP), and diastolic blood pressure (DBP) measurements from the Boppli with IAL reference values by calculating average values of mean average error (MAE) and standard deviation (SD) for each patient, then averaging those means. Safety and usability were evaluated by analysis of adverse events and survey data, respectively.
resultsThe Boppli device demonstrated good performance, meeting the FDA requirements of MAE and SD of the entire cohort: MAE (SD) 0.7 (5.3) mm Hg for MAP, -0.8 (7.7) mm Hg for SBP, and 1.4 (4.7) mm Hg for DBP. Patients with elevated MAPs, Asian ethnicity, and lower extremity IALs were the subgroups with MAE >±5 mm Hg. Various subgroups had SDs >8 mm Hg attributed to low sample sizes. The device received high usability scores from clinicians and parents. No serious adverse events were reported.
conclusionThe Boppli device is a promising alternative for continuous noninvasive BP monitoring in neonates, offering good accuracy and usability. The device, which received 510(k) clearance in September 2023, was well received by clinicians and parents, with a low-risk profile.