ArticleNMR in biomedicine2025

Improving the Accuracy and Repeatability of Renal Stiffness Measurement With Magnetic Resonance Elastography: Influence of Vibration Frequency, Acquisition Orientation, and Postprocessing-A Phantom and Volunteer Study.

Eva Jambon, Camille Courtine, Bruno Soulabaille, David Chéchin, Benjamin Robert, Joffrey Sarrazin, Gaëlle Margue, Jean-Christophe Bernhard, Yann Le Bras, Amandine Crombé

Abstract read
In one paragraph

Article in NMR in biomedicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Eva JambonDepartment of Radiology, Pellegrin Hospital, Bordeaux, France.
Camille CourtineCentre de Recherche Paul Pascal, University of Bordeaux, Pessac, France.
Bruno SoulabailleDepartment of Radiology, Saint-André Hospital, Bordeaux, France.
David ChéchinPhilips Healthcare France, Suresnes, France.
Benjamin RobertPhilips Healthcare France, Suresnes, France.
Joffrey SarrazinDepartment of Additive Fabrication Engineering, Bordeaux University Technological Institute, Gradignan, France.
Gaëlle MargueI.CaRe Team, Bordeaux Research Institute in Oncology (BRIC) INSERM U1312 and University of Bordeaux, Bordeaux, France.
Jean-Christophe BernhardI.CaRe Team, Bordeaux Research Institute in Oncology (BRIC) INSERM U1312 and University of Bordeaux, Bordeaux, France.
Yann Le BrasDepartment of Radiology, Pellegrin Hospital, Bordeaux, France.
Amandine CrombéDepartment of Radiology, Pellegrin Hospital, Bordeaux, France.ORCID https://orcid.org/0000-0003-0098-6482

Funding

ANR-21-RHUS-0015
6 · The paper itself

Abstract

MR elastography (MRE) is validated for liver fibrosis but remains highly challenging for renal imaging. Our aim was to optimize a prototype MRE sequence developed by Philips Healthcare for accurate and repeatable stiffness measurement using standardized phantoms and pipeline and to translate it to in vivo kidney imaging. Hence, in this prospective single-center study, 21 silicone-based phantoms with known stiffness ranging between 2 and 57.7 kPa were imaged at 1.5-Tesla using a 2D-gradient echo MRE acquisition. Mechanical vibrations were generated through a pneumatic driver with an actuator fixed on a dedicated 3D-printed phantom container. Four frequencies (60, 80, 100, and 120 Hz), three orientations (axial, sagittal, and coronal), and three MRE postprocessing algorithms were investigated (local frequency estimation [LFE], multifrequency dual elasto-visco inversion [MDEV], and wavenumber-based MDEV [kMDEV]). Acquisitions were repeated twice and coregistered. The best combinations of frequency, orientation, and postprocessing in terms of repeatability (measured with intraclass correlation coefficient [ICC]) and accuracy (measured with using root mean square error [RMSE]) were translated on five healthy volunteer's kidneys. The lowest/best RMSE was obtained with sagittal orientation, 120 Hz and LFE (RMSE = 8.9). The highest ICC was obtained with axial orientation, 120 Hz and LFE (ICC = 0.864, p < 0.0001). On phantoms with stiffness < 25 kPa, the best performing combination was sagittal orientation, 60 Hz and LFE (RMSE = 4.2). Multiple-way analysis of variance showed a strong influence of frequency (F = 47.8, p < 0.0001) and postprocessing (F = 49.5, p < 0.0001), but not orientation (p = 0.4771). Best image quality in volunteers was obtained with coronal orientation and mechanical vibrations of 60 Hz with anterior-posterior direction and LFE. In conclusion, our study showed that vibration frequency of 120 Hz provided the highest MRE accuracy across all phantoms (2-57.7 kPa), while 60 Hz was more suitable for stiffness values < 25 kPa and in vivo. Postprocessing significantly influenced stiffness values, with LFE offering superior accuracy and repeatability, whereas the orientation had minimal effect.

Indexed as

Elasticity Imaging TechniquesImage Processing, Computer-AssistedKidneyMagnetic Resonance ImagingPhantoms, ImagingVibrationAdultAlgorithmsFemaleHumansMaleReproducibility of Resultskidney imagingmagnetic resonance elastographyphantom studysequence optimization