Evaluation of the Characteristics and Machine Learning-Based Identification Accuracy of the International Electrotechnical Commission (IEC) 60601-1-8 Medical Equipment Auditory Alarms.
Article in Cureus, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Introduction Auditory alarms in clinical settings signal sudden changes in a patient's condition and failures in medical equipment. However, distinguishing between simultaneously sounding alarms, particularly when superimposed with various ambient sounds, remains challenging. This study aimed to develop a machine learning (ML) model for identifying auditory alarms issued by medical equipment. Methods We targeted old and new auditory alarms for medical equipment as specified in the International Electrotechnical Commission (IEC) 60601-1-8 standard. First, we evaluated the characteristics of both normal and degraded auditory alarms using cosine similarity among old and new alarms. Next, we evaluated the accuracy of ML-based identification of deteriorated alarm sound sources in both the old and new alarm groups. Results The cosine similarity among old alarms was over 0.99, while new alarms ranged from 0.886 to 0.985, and exhibited more distinct characteristics. When noise was superimposed, the similarity among old alarms increased further, making differentiation more difficult. In contrast, for most new alarms, cosine similarity values exceeded 0.99 but retained slight acoustic differences even after noise-induced degradation, demonstrating improved distinguishability. The accuracy for identifying a single degraded alarm sound was 71.9% for the support vector machine. The models exhibited a high number of misclassifications when identifying the old alarms. Conversely, the models achieved higher accuracy when classifying new alarms, with recall exceeding 80%, precision above 70%, and F-measure greater than 80% for all new alarms. The identification accuracies for two simultaneous alarms were under 20% and approximately 50% for old and new alarms, respectively. The accuracy declined when estimating two simultaneous new alarms; however, when at least one of the two alarms was correctly classified, the accuracy exceeded 90%. Conclusions This study evaluated the characteristics of old and new auditory alarms issued by medical equipment as specified in IEC 60601-1-8 and constructed ML models for identifying the type of alarms.