ArticleEuropean radiology experimental2025
Inter-phantom variability in digital mammography: implications for quality control.
Article in European radiology experimental, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Phantom studies in medical imaging (PSMI): a guide with recommendations and checklist.European radiology experimental · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPhantoms play a critical role in mammography quality control (QC) by providing standardized conditions for evaluating image quality (IQ) metrics. However, inter-phantom variability may affect the reliability of these metrics, especially for inter-system comparisons. The aim of this study was to quantify the intra- and inter-phantom variability of IQ metrics using a set of theoretically identical phantoms.
methodsTwenty-four TORMAS phantoms were imaged ten times each using a mammography unit under standardized high-dose conditions. Images were analyzed using automated software to extract 64 IQ metrics, including contrast-to-noise ratio (CNR) as well as modulation transfer function (MTF)-related and other metrics. Outliers were identified and excluded. Variability was assessed by calculating intra- and inter-phantom variances and coefficients of variation (COVs). The relative contributions of intra- and inter-phantom variability to total variability were also determined.
resultsTwo defective phantoms were excluded. Analysis of 64 IQ metrics across 22 phantoms showed higher inter-phantom variability compared to intra-phantom variability. Mean intra- and inter-phantom COVs were 6.9% and 15.1% for the 34 CNR metrics, 4.8% and 5.4% for the 5 MTF-related metrics, 0.14% and 0.75% for the 10 contrast metrics, 4.9% and 14.8% for the 15 noise metrics, respectively. Inter-phantom variability contributed 84.2% to total variability, highlighting its dominance.
conclusionInter-phantom variability significantly affects IQ metrics, emphasizing the importance of using the same phantom for inter-system comparisons to avoid confounding results. Conversely, phantoms are well-suited for assessing system reproducibility over time, focus on inter-system variability while consistently using a single phantom. RELEVANCE STATEMENT: This study highlights the significant impact of inter-phantom variability on image quality assessment, emphasizing the importance of using the same phantom for benchmarking imaging systems. These findings are crucial for optimizing quality control protocols and ensuring reliable, reproducible evaluations. KEY POINTS: Inter-phantom variability exceeded intra-phantom variability across all image quality metrics of digital mammography. Subtle details showed higher total variability compared to more distinct features. Modulation transfer function metrics exhibited comparable intra- and inter-phantom variability, highlighting positioning sensitivity. Inter-phantom variability contributes 84% to total variability, impacting imaging system comparisons. Using the same phantom ensures reliability in imaging system performance evaluations.