ArticleBioengineering (Basel, Switzerland)2025
Comparing the Efficiency of Valved Trocar Cannulas for Pars Plana Vitrectomy.
Article in Bioengineering (Basel, Switzerland), 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Hypotony-Free Closure of Infusion Sclerotomy Using a Slit-Modified Trocar in 23-Gauge Vitrectomy for Proliferative Diabetic Retinopathy.Bioengineering (Basel, Switzerland) · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo compare the efficiency of different manufacturers' valved cannulas (23G, 25G and 27G) (Alcon, Bausch & Lomb, BVI, DORC, Optikon) in maintaining intraocular pressure during vitrectomy by measuring leak pressure and the difference between set and actual intraocular pressure, under BSS and air infusion.
methodsA BSS-filled reservoir was connected to a model eye allowing placement of leak-proof valved cannulas. A pressure sensor was interposed and the bottle height increased until leakage occurred. Air leakage was measured by connecting an air pump to different manufacturers' valved cannulas, inserted upside down to blow air against the valve with inside-out direction and immersed in soapy water to detect air bubbling.
resultsThe average BSS leaking pressure was 7.69 ± 0.77 mmHg for 23G, 9.92 ± 0.57 mmHg for 25G and 7.57 ± 0.80 mmHg for 27G. The 25G valved cannulas opened at higher pressure (
conclusionDespite significant differences, all tested valved cannulas satisfy safety criteria by keeping a surgically negligible difference between the set and actual intraocular pressure. The minimal leakage measure may act as a safety pressure damper under critical conditions.