ArticleJournal of orthopaedic surgery and research2025
A safe technical modification and mid-term results in the removal of locking screws with damaged heads.
Article in Journal of orthopaedic surgery and research, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesThis study aims to describe a safe method, which does not require additional instruments, for the removal of titanium locking screws with stripped or cold welded heads. Additionally, it evaluates the outcomes of patients in whom this method was applied.
methodsThe proposed method involves using a 3.0 ball-type abrader attached to a standard orthopedic machine to abrade the screw head until it completely separates from the plate hole. This technique was compared to the conventional high-speed burr method, focusing on the maximum temperature generated during the process, as measured with a thermal camera on an artificial bone model. Additionally, the outcomes of 22 patients treated with this technique and followed for at least 12 months were evaluated.
resultsThe surface temperature of the plate reached 101.4 °C during the high-speed burr technique. In contrast, when the same technique was performed using the orthopedic machine, the plate surface temperature increased to 62.4 °C. This technique was applied to 22 patients, with an average age of 37 years, and a mean follow-up period of 18.5 months. No complications were reported in any of the patients. Among the removed screws, 14 (63.7%) were 3.5 mm, 4 (18.2%) were 2.7 mm, 3 (13.6%) were 2.4 mm, and 1 (4.5%) was 4.5 mm. Cold welded was observed in only one 4.5 mm screw, while head stripping was observed in all other cases.
conclusionThis modified technique, using a standard orthopedic machine and burr tip, provides a safe and effective method for removing damaged screws that does not require complex or additional instruments.