Ultrashort-T2* mapping at 7 tesla using an optimized pointwise encoding time reduction with radial acquisition (PETRA) sequence at standard and extended echo times.
Article in PloS one, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Article
Corrections and comments
- Erratum issued
Authors and funding
3 authors.
Funding
Abstract
Zero echo time (ZTE) sequences capture signal from tissues with extremely short T2* and are useful for qualitative and quantitative imaging of musculoskeletal tissues' ultrashort-T2* components. One such sequence is Pointwise Encoding Time Reduction with Radial Acquisition (PETRA). While this sequence has shown promising results, it has undergone only limited testing at 7 tesla (T). The purpose of this work was to evaluate PETRA at 7T in its standard, commercially available form and with sequence code modifications to allow extended echo times for the purpose of performing ultrashort-T2* mapping. We acquired PETRA images of MnCl2 and collagen phantoms and of the knee in eight participants (5 for optimization and 3 for ultrashort-T2* mapping assessment; 5 male/3 female, 39 ± 11 years old). Images were acquired using a 1-transmit/28-receive-channel knee coil. Artifacts, signal, signal-to-noise ratio (SNR), ultrashort-T2*, the corresponding curve fit quality, and repeatability were assessed. In knee tissues, SNR was higher at TE = 0.07 msec than in a conventional-TE sequence (Dual-Echo Steady State with TE = 2.55 msec), with values of 68-337 for PETRA versus 16-30 for the same regions in the conventional-TE series. Acquisition of series for ultrashort-T2* maps was feasible at 1.50 mm isotropic acquisition resolution and TE ≤ 0.58 msec. Strong linear correlations were observed between relaxation rates (R2*) and MnCl2 concentration, and between signal and collagen concentration. Ultrashort-T2* signal decay curve fit R2 and repeatability were high for phantom and knee ultrashort-T2* <1 msec. PETRA imaging with minimal artifacts, high SNR, and scan time < 11 minutes was achieved at 7T at high (0.34 mm isotropic) resolution at TE = 0.07 msec and lower resolution (1.52 mm isotropic) at echo times ≤ 0.58 msec. Ultrashort-T2* mapping provided acceptable curve-fitting results for substances with sub-millisecond T2*.