ArticleInternational journal of laboratory hematology2025
Automated Red and White Blood Cell Counting in Capillary Microsamples by Sysmex-XN Predilution Mode.
Article in International journal of laboratory hematology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Integrated ESR and PLT-H Measurement Using the BC-6800 Plus Hematology Analyzer: A Comprehensive Analytical Evaluation.Journal of clinical laboratory analysis · 2026Article
- Nano-Enabled Microfluidic Platforms for Functional Immunomonitoring in Pediatric Sepsis.International journal of nanomedicine · 2026Review
- Automated Red and White Blood Cell Counting in Capillary Microsamples by Sysmex-XN Predilution Mode.International journal of laboratory hematology · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesNeonatal patients present a challenge to the clinical laboratory because of their low blood volume. The Sysmex XN-series features a predilution (PD) mode allowing hematologic measurements with only 20-50 μL of blood. We verified the PD mode for analysis of selected hematologic parameters in 50 μL microsamples.
methodsMicrosamples were prepared from adult EDTA blood. White blood cell count (leukocytes, neutrophils, and lymphocytes) and red blood cell parameters (erythrocytes, hemoglobin, hematocrit, mean cell volume, and reticulocytes) were evaluated. The imprecision of the PD mode was evaluated over 3 days, and the accuracy was assessed by method comparison with the standard whole blood mode. The effect of capillary sampling, microsample stability during storage, and pneumatic tube transportation was evaluated. Finally, the bias was reproduced in a small sample of pediatric patients.
resultsFor white blood cells, the bias was ≤ 5.4% (95% CI: 4.5-6.2) and imprecision was below 3.5% (except at the lowest levels). Capillary sampling had little effect on PD analytical performance (bias ≤ 0.7% and imprecision ≤ 4.7%) and white blood cell counts were stable for 7 h at room temperature and after pneumatic tube transportation. The red blood cell parameters generally exceeded the allowable bias and imprecision. The bias of the pediatric samples remained within the 95% PI for the method comparison studies.
conclusionThe PD mode has acceptable analytical performance and preanalytical stability for white blood cell counts but not for red blood cell parameters. It may offer a low-volume alternative for hematologic monitoring.