Sensing and detection performance of the novel, small-diameter OmniaSecure defibrillation lead: in-depth analysis from the LEADR trial.
Article in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT07406737 (Feasibility and Outcomes Registry of Lumenless dEfibrillation Leads for Conduction System Pacing), which is not on this map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Feasibility and Outcomes Registry of Lumenless dEfibrillation Leads for Conduction System Pacing
Who cites it
2 citing papers in PubMed.
- Safety and Efficacy of the Novel OmniaSecure Defibrillation Lead Through Long-Term Follow-Up: Final Results From the LEADR Trial.Circulation. Arrhythmia and electrophysiology · 2026Trial
- Controversy: in heart failure patients with a reduced ejection fraction and left bundle branch block, conduction system pacing can be a valid alternative to biventricular pacing-pro and contra.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
21 authors.
Funding
Abstract
aimsThe Lead EvaluAtion for Defibrillation and Reliability (LEADR) trial evaluated the small-diameter (4.7 Fr), integrated bipolar OmniaSecure defibrillation lead. As previously reported, the trial exceeded primary safety and efficacy objective thresholds, demonstrating favourable performance and zero fractures through ∼12 months follow-up, with patients in ongoing follow-up. Longer-term follow-up of the LEADR trial with emphasis on the sensing and detection capabilities of the OmniaSecure lead is reported here. METHODS AND
resultsPatients with indications for de novo implantable cardioverter-defibrillators/cardiac resynchronisation therapy defibrillators were implanted with the OmniaSecure lead in standard right ventricle (RV) locations and followed at pre-specified intervals along with CareLink™ remote monitoring transmissions, where available. Throughout follow-up, the lead was evaluated for safety, efficacy, and reliability along with sensing and detection performance. There were 643/657 (97.9%) patients successfully implanted with the OmniaSecure lead with mean follow-up of 18.2 ± 5.5 months. There was a 96.9% freedom from major study lead-related complications at 24 months. Inappropriate shock rate was 2.7 and 3.8% at 12 and 24 months, respectively. At 24 months, 17.6% of patients received appropriate therapies (shock and/or ATP) with a 76.5% ATP efficacy. There have been zero fractures during follow-up along with chronically stable pacing capture threshold, pacing impedance, and R-wave amplitudes. There were four patients with an adverse event related to PWOS (0.6%), none of which was associated with inappropriate shock. There were four patients with an adverse event related to TWOS (0.6%), of which three patients were associated with inappropriate shock (0.5%). Oversensing was resolved predominantly by programming the RV sensitivity to less sensitive settings. During VF induction at implant, 97.6% (120/123) of patients showed appropriate VF episode detection at the least sensitive setting of 1.2 mV, with the remaining having detection at more sensitive settings. In follow-up, 670 VT/VF episodes were appropriately detected and treated in 94 patients with a variety of RV sensitivities and no reports of under-detected episodes. Moreover, a virtual sensitivity analysis also showed no under-detection across different RV sensitivity programming.
conclusionChronic sensing performance of the OmniaSecure defibrillation lead demonstrated R-wave stability with a low rate of P-wave and T-wave oversensing, resolved predominantly by adjusting RV sensitivity. Further, VT/VF detection was successful and was not impacted when programmed to less sensitive settings. The OmniaSecure lead shows robust sensing and detection performance and programmability in ongoing follow-up.