Evidence map›Paper›PMID 40127675›Full record

ArticleEuropace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology2025

Sensing and detection performance of the novel, small-diameter OmniaSecure defibrillation lead: in-depth analysis from the LEADR trial.

Prashanthan Sanders, Pamela K Mason, Bert Hansky, Paolo De Filippo, Maully J Shah, Darius P Sholevar, John S Zakaib, Francois Philippon, Bernice Tsang, Rajeev K Pathak and 11 more

Registry-linked trialAbstract readAdaptive Clinical TrialMulticenter Study
In one paragraph

Article in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT07406737 (Feasibility and Outcomes Registry of Lumenless dEfibrillation Leads for Conduction System Pacing), which is not on this map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT07406737 not yet recruitingnot on this mapstarted 2026, after this paper: background citation

Feasibility and Outcomes Registry of Lumenless dEfibrillation Leads for Conduction System Pacing

Typeobservational_patient_registrySponsorUniversity Hospital, CaenRan2026 to 2027Enrolled200ConditionsCardiac Resynchronization Therapy, Left Bundle Branch Area Pacing
3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Trial
  2. Controversy: in heart failure patients with a reduced ejection fraction and left bundle branch block, conduction system pacing can be a valid alternative to biventricular pacing-pro and contra.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2025
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

21 authors.

Prashanthan SandersCentre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, South Australia, Australia.ORCID 0000-0003-3803-8429
Pamela K MasonElectrophysiology Laboratory, University of Virginia, Charlottesville, VA, USA.ORCID 0000-0002-4276-2391
Bert HanskyKlink für Kardiologie und Internistische Intensivmedizin, Städtische Kliniken, Bielefeld, Germany.ORCID 0009-0002-4592-7734
Paolo De FilippoCardiac Electrophysiology and Pacing, ASST Papa Giovanni XXIII, Bergamo, Italy.ORCID 0000-0002-1144-5000
Maully J ShahCardiac Center, The Children's Hospital, Philadelphia, PA, USA.
Darius P SholevarCardiology Group, Virtua Health, Cherry Hill, NJ, USA.
John S ZakaibHeart Rhythm Center, Minneapolis Heart Institute Foundation, Minneapolis, MN, USA.ORCID 0009-0001-3754-6555
Francois PhilipponInstitut Universitaire de Cardiologie et de Pneumologie de Québec, Laval University, Québec City, Québec, Canada.ORCID 0000-0002-2012-0486
Bernice TsangCardiac Program, Southlake Regional Health Centre, Newmarket, Ontario, Canada.ORCID 0009-0000-4081-1033
Rajeev K PathakCardiac Electrophysiology & Pacing, Canberra Heart Rhythm Centre and Australian National University, Garran, ACT, Australia.
Travis D RichardsonDivsion of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, TN, USA.ORCID 0000-0002-9772-0308
Meir FriedmanHeart & Vascular Institute Wellness Center, Hartford Hospital, Hartford, CT, USA.
Robert D SchallerCardiac Arrhythmia Program, Hospital of the University of Pennsylvania, Philadelphia, PA, USA.ORCID 0000-0001-8608-7247
Ignasi AngueraUnidad de Arritmias y Electrofisiologia, Hospital Universitari Bellvitge, Barcelona, Spain.ORCID 0000-0001-6988-3487
Attila MihalczKlinische Abteilung für Innere Medizin 1, Universitätsklinikum Krems, Krems, Austria.
Babak BozorgniaCardiology Department, Lehigh Valley Hospital-Cedar Crest, Allentown, PA, USA.
Amy E ThompsonCardiac Rhythm Management, Medtronic, Inc., Minneapolis, MN, USA.ORCID 0000-0002-7363-5870
Katherin AriasCardiac Rhythm Management, Medtronic, Inc., Minneapolis, MN, USA.
Baerbel MausCardiac Rhythn Management, Medtronic Bakken Research Center, Maastricht, The Netherlands.
Chad BoundsCardiac Rhythm Management, Medtronic, Inc., Minneapolis, MN, USA.
George H CrossleyDivsion of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, TN, USA.ORCID 0000-0003-0096-8401

Funding

Medtronic Inc.
6 · The paper itself

Abstract

aimsThe Lead EvaluAtion for Defibrillation and Reliability (LEADR) trial evaluated the small-diameter (4.7 Fr), integrated bipolar OmniaSecure defibrillation lead. As previously reported, the trial exceeded primary safety and efficacy objective thresholds, demonstrating favourable performance and zero fractures through ∼12 months follow-up, with patients in ongoing follow-up. Longer-term follow-up of the LEADR trial with emphasis on the sensing and detection capabilities of the OmniaSecure lead is reported here. METHODS AND

resultsPatients with indications for de novo implantable cardioverter-defibrillators/cardiac resynchronisation therapy defibrillators were implanted with the OmniaSecure lead in standard right ventricle (RV) locations and followed at pre-specified intervals along with CareLink™ remote monitoring transmissions, where available. Throughout follow-up, the lead was evaluated for safety, efficacy, and reliability along with sensing and detection performance. There were 643/657 (97.9%) patients successfully implanted with the OmniaSecure lead with mean follow-up of 18.2 ± 5.5 months. There was a 96.9% freedom from major study lead-related complications at 24 months. Inappropriate shock rate was 2.7 and 3.8% at 12 and 24 months, respectively. At 24 months, 17.6% of patients received appropriate therapies (shock and/or ATP) with a 76.5% ATP efficacy. There have been zero fractures during follow-up along with chronically stable pacing capture threshold, pacing impedance, and R-wave amplitudes. There were four patients with an adverse event related to PWOS (0.6%), none of which was associated with inappropriate shock. There were four patients with an adverse event related to TWOS (0.6%), of which three patients were associated with inappropriate shock (0.5%). Oversensing was resolved predominantly by programming the RV sensitivity to less sensitive settings. During VF induction at implant, 97.6% (120/123) of patients showed appropriate VF episode detection at the least sensitive setting of 1.2 mV, with the remaining having detection at more sensitive settings. In follow-up, 670 VT/VF episodes were appropriately detected and treated in 94 patients with a variety of RV sensitivities and no reports of under-detected episodes. Moreover, a virtual sensitivity analysis also showed no under-detection across different RV sensitivity programming.

conclusionChronic sensing performance of the OmniaSecure defibrillation lead demonstrated R-wave stability with a low rate of P-wave and T-wave oversensing, resolved predominantly by adjusting RV sensitivity. Further, VT/VF detection was successful and was not impacted when programmed to less sensitive settings. The OmniaSecure lead shows robust sensing and detection performance and programmability in ongoing follow-up.

Indexed as

Cardiac Resynchronization TherapyCardiac Resynchronization Therapy DevicesDefibrillators, ImplantableElectric CountershockAgedElectrodes, ImplantedEquipment DesignFemaleHumansMaleMiddle AgedReproducibility of ResultsTime FactorsTreatment OutcomeAnti-tachycardia pacingImplantable cardioverter-defibrillatorInappropriate shocksLumenless leadsP-wave oversensing

Identifiers

PMID40127675
PMCPMC12001238

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials