Evidence map›Paper›PMID 40096052›Full record

ArticleSensors (Basel, Switzerland)2025

Reduced Heating Wireless Energy Transmission System for Powering Implanted Circulatory Assist Devices: Benchtop and In-Vivo Studies.

Mohammad L Karim, Rachel Grimes, Harry Larkin, Antonio M Bosnjak, James McLaughlin, Paul Crawford, David McEneaney, Omar J Escalona

Abstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Mohammad L KarimNanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.ORCID 0000-0002-0650-5232
Rachel GrimesNanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.
Harry LarkinNanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.
Antonio M BosnjakNanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.
James McLaughlinNanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.ORCID 0000-0001-6026-8971
Paul CrawfordPaul Crawford Veterinary Services, Larne BT40 3RW, UK.ORCID 0000-0003-3666-6050
David McEneaneyGalvani TECH Ltd., School of Engineering, Ulster University, Belfast BT15 1AP, UK.
Omar J EscalonaNanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.ORCID 0000-0001-6696-5911

Funding

Innovate UK 10078920
6 · The paper itself

Abstract

This study aimed to develop a novel Transdermal Energy Transmission System (TETS) device that addresses the driveline complications faced by patients with advanced heart failure (HF). Our TETS device utilizes a two-channel configuration with a very-low duty cycle and a pulsed RF power transmission technique, along with elliptically shaped flexible coil inductive coupling elements. We integrated a battery charging controller module into the TETS, enabling it to recharge an implanted Lithium-Ion (Li-Ion) battery that powers low-power-rated Circulatory Assist Devices, or left ventricular assist devices (LVADs). Benchtop measurements demonstrated that the TETS delivered energy from the implanted coils to the battery charging module, at a charging rate of up to 2900 J/h, presented an average temperature increase (ΔT) of 3 °C. We conducted in vivo measurements using four porcine models followed by histopathological analysis of the skin tissue in the implanted coils areas. The thermal profile analysis from the in vivo measurements and the calculated charging rates from the current and voltage waveforms, in porcine models, indicated that the charging rate and temperature varied for each model. The maximum energy charging rate observed was 2200 J/h, with an average ΔT of 3 °C. The exposed skin tissue histopathological analysis results showed no evidence of tissue thermal damage in the in vivo measurements. These results demonstrate the feasibility of our developed TETS device for wireless driving implanted low-power-rated LVADs and Li-Ion charging.

Indexed as

Electric Power SuppliesHeart-Assist DevicesWireless TechnologyAnimalsHeart FailureHumansLithiumSwineLithiumleft ventricular assist device (LVAD)rechargeable implanted medical batterytissue heating effectstranscutaneous energy transmission system (TETS)wireless charging rate

Identifiers

PMID40096052
PMCPMC11902738

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.