ArticleJournal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine2025
The influence of an innovative inspiratory pressure relief accessory on continuous positive airway pressure profiles in a laboratory bench setting.
Article in Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Safety principles of lowering inspiratory pressure.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
STUDY
objectivesDespite its efficacy, long-term adherence to continuous positive airway pressure (CPAP) hovers around 50%. The V-Com (SleepRes, Murfreesboro, TN), an inspiratory pressure relief (IPR) accessory, claims to improve comfort by decreasing inspiratory pressures. We aim to independently assess its impact on fixed CPAP settings in a laboratory bench model.
methodsAn ASL-5000 Breathing simulator (IngMar Medical, Pittsburgh, PA) with normal lung settings was employed. A ResMed AirSense 11 device was connected to ASL-5000 through a React Health Rio II exhalation vent and an elbow adapter. Mean inspiratory and expiratory pressures were measured at fixed CPAP settings (4, 6, 8, 10, 12, 14, 16, 18, and 20 cm H
resultsWith 1 IPR, mean inspiratory pressures decreased significantly from the baseline pressures (
conclusionsThis IPR accessory decreases inspiratory pressures but only marginally reduces expiratory pressures in this laboratory bench model. Two IPR units further decrease both inspiratory and expiratory pressures. CITATION: Sleeper GP, Catullo KA, Chen P-L, Patil SP, Strohl KP, Chiang AA. The influence of an innovative inspiratory pressure relief accessory on continuous positive airway pressure profiles in a laboratory bench setting.