Safety and feasibility of a smart assistive bone-cement injection system: a cadaveric study.

Chen Jin, Ming-Liang Ning, Rui-Jun Xu, Xiao-Jian Ye, Hao-Jie Chen, Jiang-Ming Yu

Abstract read
In one paragraph

Article in Journal of orthopaedic surgery and research, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Chen Jin *Department of Orthopedics, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, 1111 Xianxia Road, Changning District, Shanghai, 200050, China.
Ming-Liang Ning *Department of Orthopedics, Changzhou Geriatric hospital affiliated to Soochow University, Changzhou, 213000, China.
Rui-Jun XuDepartment of Orthopedics, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, 1111 Xianxia Road, Changning District, Shanghai, 200050, China.
Xiao-Jian YeDepartment of Orthopedics, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, 1111 Xianxia Road, Changning District, Shanghai, 200050, China.
Hao-Jie ChenDepartment of Orthopedics, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, 1111 Xianxia Road, Changning District, Shanghai, 200050, China. chenhaojie97@126.com.
Jiang-Ming YuDepartment of Orthopedics, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, 1111 Xianxia Road, Changning District, Shanghai, 200050, China. yjm_st@163.com.

Funding

Laboratory Open Fund of Key Technology and Materials in Minimally Invasive Spine Surgery of Shanghai Tongren Hospital 2024JZWC-YBB02Medical and Industrial cross research Foundation of "Star of Jiaotong University" Program of Shanghai Jiao Tong University, China YG2022ZD030National Natural Science Foundation of China 82401609Science and Technology Commission of Shanghai Municipality 24ZR1463700Scientific Research Staring Foundation of Shanghai Tongren Hospital TR2023rc10
6 · The paper itself

Abstract

backgroundPercutaneous vertebral augmentation is an effective and commonly surgical treatment for osteoporotic vertebral compression fractures, but the problem of bone cement leakage still cannot be prevented. It has been reported that cement leakage occurs in approximately 20% of vertebroplasty procedures, with symptomatic manifestations reported in 1.6% of cases. Leakage of bone cement into the spinal canal increases the risk for spinal cord compression and nerve injury. The objective of this study was to introduce a smart assistive device specifically designed to facilitate both cement injection control and operator protection.

methodsTwo freshly frozen human cadaver specimens were used. The 2 cadaver specimens were divided according to injection method: manual (10 vertebrae, T8-L5); and motorized (10 vertebrae, T8-L5). Fluoroscopy time, cement time, volume injected, and cement distribution were recorded. Postoperative radiography and CT images were used to assess cement distribution in this cadaveric study.

resultsThe number of times intraoperative X-ray fluoroscopy was used for the manual injection group (6.7 ± 1.5) was significantly greater (P < 0.001) than that for the motorized injection group (4.1 ± 0.9). Mean cement time for the manual injection group (164.3 ± 18.7 s) was significantly greater (P < 0.001) than that for the motorized injection group (72.0 ± 7.2 s). There were no significant differences in the amount of cement injected in the manual vs. motorized injection group (5.2 ± 1.3 mL vs. 5.3 ± 1.0 mL; P = 0.878). Moreover, we found that leakage of cement outside the vertebral body was noted in 4 of 10 injected vertebrae (40%) in the manual injection group, whereas there was no bone cement leakage in the motorized injection group.

conclusionsThe system exhibited more precise control of the bone cement injection dosage and better cement distribution compared with traditional manual injection. In addition, the device provided remote activation, reducing the X-ray intake of the surgeon.

Indexed as

Bone CementsSpinal FracturesVertebroplastyAgedCadaverFeasibility StudiesFemaleFluoroscopyHumansInjectionsLumbar VertebraeMaleThoracic VertebraeBone CementsBiomedical equipmentCadaveric trialPercutaneous vertebral augmentationRemote handlingSmart assistive

Identifiers

PMID40075404
PMCPMC11905486

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.