ArticleIEEE transactions on bio-medical engineering2025
Time of Flight Transmission Mode Ultrasound Computed Tomography With Expected Gradient and Boundary Optimization.
Article in IEEE transactions on bio-medical engineering, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
objectiveQuantitative time of flight in transmission mode ultrasound computed tomography (TFTM USCT) is a promising, cost-effective, and non-invasive modality, particularly suited for functional imaging. However, TFTM USCT encounters resolution challenges due to path information concentration in specific medium regions and uncertainty in transducer positioning. This study proposes a method to enhance resolution and robustness, focusing on low-frequency TFTM USCT for pulmonary imaging.
methodsThe proposed technique improves the orientation of steepest descent algorithm steps, preventing resolution degradation due to path information concentration, while allowing for a posteriori sensor positioning retrieval. Total variation regularization is employed to stabilize the inverse problem, and a modified Barzilai-Borwein method determined the step size in the steepest descent algorithm. The proposed method was validated through simulations of data on healthy and abnormal cross-sections of a human chest using MATLAB's k-Wave toolbox. Additionally, experimental data were collected using a Verasonics Vantage 64 low-frequency system and a ballistic gel torso-mimicking phantom to assess robustness under a more realistic environment, closer to that of a clinical situation.
resultsThe results showed that the proposed method significantly improved image quality and successfully retrieved sensor locations from imprecise positioning. SIGNIFICANCE: This study is the first to address transducer location uncertainty on a transducer belt in TFTM USCT and to apply an estimated gradient approach. Additionally, low-frequency USCT for lung imaging is quite novel, and this work addresses practical questions that will be important for translational development.
Indexed as