Performance evaluation of Rayto RT-7600Vet hematology analyzer in side-by-side comparison with manual hematological methods for apparently healthy Cholistani cattle blood.

Umer Farooq, Mushtaq Hussain Lashari, Zia Ur Rehman, Musadiq Idris, Haroon Rashid, Shagufta Nasreen, Farah Laraib, Rubaisha Ameer, Maryam Chauhdary, Iram Fatima

Abstract readComparative Study
In one paragraph

Article in PloS one, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Umer FarooqDepartment of Physiology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.ORCID 0000-0002-2708-5599
Mushtaq Hussain LashariDepartment of Zoology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Zia Ur RehmanDepartment of Physiology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Musadiq IdrisDepartment of Physiology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Haroon RashidDepartment of Physiology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Shagufta NasreenDepartment of Zoology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Farah LaraibDepartment of Zoology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Rubaisha AmeerDepartment of Zoology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Maryam ChauhdaryDepartment of Zoology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.
Iram FatimaDepartment of Zoology, The Islamia University of Bahawalpur, Bahawalpur, Pakistan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The present study is the first from Pakistan being reported with an objective to assess performance of Rayto RT-7600Vet hematology analyzer (HA) for Cholistani cattle blood (n = 134), in comparison to the manual hematological methods. The four hematological attributes viz. total erythrocyte count (TEC), hemoglobin (Hb), packed cell volume (PCV) and platelet count (PLT) were deduced through HA (A) and manual (M) methods. Various statistical tests were implied to ascertain level of interrelationship, accuracy and level of agreement between the two methods. All attributes attained through manual methods had high positive, leptokurtic distribution (having many outliers) except for PLT-M and PCV-M. The coefficient of variation for attributes attained through HA and manual methods ranged from 16-24% and from 16-59%, respectively. Comparison between the overall results revealed that all the studied attributes, except TEC, were significantly (P≤0.05) different for both methods. A weak relationship was noticed between the attributes attained through two methods as indicated by weak r-values and adjusted r-square values. The reliability level of estimating Hb and PCV had highest intraclass correlation coefficient value of 0.722 and 0.555 for average measures, respectively. However, accuracy level, as determined through Lin's concordance correlation coefficient was highest for TEC (0.9504) for both analytical methods. Poor level of agreement, in general, was shown for the two methods of analysis regarding all four hematological attributes through Bland and Altman test. In conclusion, the Rayto RT-7600Vet) may present data having higher skewness, kurtosis, and CV%, however, they are valid for multi-species hematological analysis. Caution must however, be taken in interpreting their results with corrected reference intervals and CV% for each machine and for each tested attribute.

Indexed as

Hematologic TestsHematologyAnimalsCattleErythrocyte CountHematocritHemoglobinsPakistanPlatelet CountReproducibility of ResultsHemoglobins

Identifiers

PMID40067785
PMCPMC11896046

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.