Field output correction factors using a scintillation detector.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Characterization of a novel plastic scintillation detector for in vivo electron dosimetry.Journal of applied clinical medical physics · 2026Article
- Comprehensive characterization and validation of a fast-resolving (1000 Hz) plastic scintillator for ultra-high dose rate electron dosimetry.Medical physics · 2025Article
- Field output correction factors using a scintillation detector.Medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
Abstract
backgroundAs small radiation fields play an ever-increasing role in radiation therapy, accurate dosimetry of these fields becomes critical to ensure high quality dose calculation and treatment optimization. Despite the availability of several small volume dose detectors, small field dosimetry remains challenging. The PRB-0002, a new plastic scintillation detector part of the Hyperscint RP-200 dosimetric platform from Medscint, that requires only minimal corrections can potentially facilitate small field measurements. PURPOSE: The main objective of this work is to adapt small field formalism to plastic scintillation dosimetry using both Monte Carlo (MC) simulations and measurements. The secondary objective is to use the fully characterized PRB-0002 for accurate and precise measurements of field output correction factors and compare those measurements with that from other small field detectors.
methodsOur work is based on IAEA TRS-483 report. EGSnrc MC simulations of the PRB-0002 were conducted to determine the impact of detector composition, surrounding materials, dose averaging within the sensitive volume as well as ionization quenching. From these simulations, the field output correction factors of PRB-0002 were determined. Then, by experimental comparisons, field output correction factors for 2 solid state detectors and 3 small volume ion chambers have been determined.
resultsWith PRB-0002, the material composition factor is well balanced with the ionization quenching making the field output correction factor near unity. For fields between 0.6
conclusionsWith field output correction factors close to unity for a wide range of field sizes, the PRB-0002 is a near-ideal detector for small field dosimetry. Furthermore, it can be used to experimentally determine the field output correction factors of other dosimeters with great accuracy.