Evidence map›Paper›PMID 40041035›Full record

ArticleEuropean heart journal. Imaging methods and practice2025

The impact of bicuspid valve morphology on the selection of transcatheter aortic valve implantation devices: an

Benedetta Grossi, Giulia Luraghi, Sara Barati, Chiara Forte, Luca Gerosa, Ottavia Cozzi, Fabrizio D'Ascenzo, Gianluigi Condorelli, Francesco Migliavacca, Giulio Stefanini

Abstract read
In one paragraph

Article in European heart journal. Imaging methods and practice, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Benedetta GrossiDepartment of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, Pieve Emanuele, Milan 20072, Italy.ORCID https://orcid.org/0009-0005-2337-5176
Giulia LuraghiDepartment of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, Milan 20133, Italy.ORCID https://orcid.org/0000-0002-0434-8544
Sara BaratiDepartment of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, Milan 20133, Italy.
Chiara ForteDepartment of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, Milan 20133, Italy.
Luca GerosaDepartment of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, Milan 20133, Italy.
Ottavia CozziDepartment of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, Pieve Emanuele, Milan 20072, Italy.ORCID https://orcid.org/0000-0002-0700-1781
Fabrizio D'AscenzoDivision of Cardiology, Cardiovascular and Thoracic Department, Azienda Ospedaliero-Universitaria Città della Salute e della Scienza di Torino, Corso Bramante 88, Turin 10126, Italy.
Gianluigi CondorelliDepartment of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, Pieve Emanuele, Milan 20072, Italy.
Francesco MigliavaccaDepartment of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, Milan 20133, Italy.ORCID https://orcid.org/0000-0003-4644-630X
Giulio StefaniniDepartment of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, Pieve Emanuele, Milan 20072, Italy.ORCID https://orcid.org/0000-0002-3558-6967

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Aims: Bicuspid aortic valve (BAV) represents a challenge for transcatheter aortic valve implantation (TAVI). Few data are reported about the procedural implications of BAV using different self-expandable devices. The aim of this study is to investigate how BAV and tricuspid aortic valve (TAV) morphologies influence device selection and their impact on the potential development of post-operative conduction disturbances, using a novel Methods and results: Five patients with BAV undergoing TAVI were enrolled. TAVs were virtually modelled within each BAV patient-specific anatomy, resulting in 10 virtual patients. Acurate Neo2 and Evolut R implantations were subsequently simulated across all cases. Post-implantation stresses exerted on both the stent and aortic root were measured, allowing a comparative analysis of the impact of the two valve morphologies. Comparing stent stresses between BAV and TAV configurations, the stress gap increased by 21.96 ± 5.35% ( Conclusion: Our patient-specific model shows that high radial force devices, such as Evolut R, demonstrate consistent expansion regardless of valve morphology, without increasing the risk of post-implantation conduction disturbances, hence resulting more suitable for BAV cases. Incorporating this methodology into pre-operative planning could support clinicians in selecting the most suitable device with a patient-specific approach.

Indexed as

bicuspid aortic valvefinite element methodpatient-specific simulationsTAVI

Identifiers

PMID40041035
PMCPMC11879518

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.