Simple but rigorous procedure for ultrasound quality assurance.

Paul L Carson, Megan K Russ, Stephen Z Pinter, Sander Dekker

Abstract read
In one paragraph

Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Paul L CarsonDepartment of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
Megan K RussDepartment of Radiology, Duke University, Durham, North Carolina, USA.
Stephen Z PinterDepartment of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
Sander DekkerCablon Medical B.V., Leusden, Utrecht, Netherlands, Netherlands.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundMedical ultrasound (US) system quality assurance (QA) is employed only modestly by users because of the large number of variables recommended for measurement and the relatively low use of all that information when it is acquired. By 2011 it was again clear that the equipment was not as reliable as often contended; that failures were mainly transducer, cable, and visible mechanical effects, easily addressed by uniformity tests and visual inspections. Monitor degradation also became more widely recognized. Recently it was proposed that system imaging performance QA could be accomplished by simple, cine loop acquisitions with a random hypoechoic sphere phantom (RHSP). With automated analysis, this should detect almost all B mode image quality difficulties and sources of most defects in other modes. The measures, based on the spheres' mean lesion signal-to-noise ratios (LSNRms), also would be absolute tests in 3D and valid for acceptance testing and comparisons, providing intuitive results for users to anticipate a transducer's utility in various depths and applications. This challenge to traditional phantoms has stimulated improvement in more traditional polymer filament or wire phantoms, making it easier to see poor slice thickness, but with considerable acquisition effort and cost. PURPOSE: To evaluate the potential and ease of use of the commercially available RHSPs, we compared them with a new wire phantom and an established uniformity phantom for detecting physically simulated, common transducer defects, using commercially available automated analysis.

methodsThe three types of transducer defects were simulated physically for comparison with the same, undisrupted, transducer. A layer of Styrofoam was taped over either one end of the array, one side of the array, one part of one side of the array, or not at all, essentially blocking all of the elements on one end or one side, or just part of the elevational extent of the involved elements. Scanning methods and analysis, generally described in IEC TS62736, provided median lesion signal-to-noise ratio (LSNRm) and clarity index (Ci) on the spheres, -6 dB pulse-echo response widths on the wires and median signal level on the uniformity phantom and RHSP. Several defective transducers were evaluated as well.

resultsThe RHSP phantom/analysis gave consistent results, with delineation of the lateral and the full elevational disruptions, but only meager delineation of the partial elevational disruption given chosen settings. Beam width measurements on the 90°/45° wire phantom delineated all three disruptions but were more difficult to perform and function over limited settings. The uniformity measurements revealed only the lateral disruption.

conclusionsThe RHSP phantom is now available at a considerably lower cost than wire-based phantoms. It demonstrated good detection of all but the subtlest disruption. We preliminarily recommend this phantom for US system user QA and performance evaluation, after initial verification on each model of the system of the grayscale characteristic curve and spatial measurements. Quantitative expectations for different US systems will develop rapidly, as more users will be implementing this simpler QA. Software or phantom modifications can add evaluation of spatial measurements.

Indexed as

Quality Assurance, Health CarePhantoms, ImagingQuality ControlSignal-To-Noise RatioUltrasonographymedical imagingperformance evaluationquality control